An optosensor for riboflavin with C-18 silica gel as a substrate

被引:21
作者
Gong, ZL
Zhang, ZJ
机构
[1] SHAANXI NORMAL UNIV,DEPT CHEM,XIAN 710062,PEOPLES R CHINA
[2] NANJING UNIV,DEPT CHEM,NANJING 210008,PEOPLES R CHINA
基金
中国国家自然科学基金;
关键词
sensors; riboflavin; optosensor; C-18 silica gel;
D O I
10.1016/S0003-2670(96)00447-3
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A flow-through optosensor for the sensitive determination of riboflavin in some pharmaceutical preparations was developed. The sensor was developed in conjunction with a flow-injection analysis system and uses C-18 silica gel as a substrate. The wavelengths of excitation and emission were 466 and 515 nm, respectively. The applicable concentration range was 2.2-91.8 ng ml(-1) with a detection limit (3 sigma-criterion) of 0.4 ng ml(-1). The relative standard deviation for ten determinations of 15 ng ml(-1) of riboflavin was 1.1%. The method is more sensitive than other methods described in the literature.
引用
收藏
页码:161 / 165
页数:5
相关论文
共 11 条
[11]  
ZHEBENTYAEV A I, 1986, Farmatsiya (Moscow), V35, P16