Nonenzymatic glucose detection using mesoporous platinum

被引:578
作者
Park, S
Chung, TD
Kim, HC
机构
[1] Sungshin Womens Univ, Dept Chem, Seoul 136742, South Korea
[2] Seoul Natl Univ, Coll Med, Dept Biomed Engn, Seoul 110744, South Korea
[3] Seoul Natl Univ, Med Res Ctr, Inst Med & Biol Engn, Seoul 110744, South Korea
关键词
D O I
10.1021/ac0263465
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Roughness of nanoscopic dimensions can be used to selectively enhance the faradaic current of a sluggish reaction. Using this principle, we constructed mesoporous structures on the surfaces of pure platinum electrodes responding even more sensitively to glucose than to common interfering species, such as L-ascorbic acid and 4-acetamidophenol. Good sensitivities, as high as 9.6 muA cm(-2) mM(-1), were reproducibly observed in the presence of high concentration of chloride ion. The selectivities, sensitivities, and stabilities determined experimentally have demonstrated the potential of mesoporous platinum as a novel candidate for nonenzymatic glucose sensors.
引用
收藏
页码:3046 / 3049
页数:4
相关论文
共 24 条