Reducing no flow times during automated external defibrillation

被引:35
作者
Eilevstjonn, J
Kramer-Johansen, J
Eftestol, T
Stavland, M
Myklebust, H
Steen, PA
机构
[1] Laerdal Med AS, N-4002 Stavanger, Norway
[2] Ullevaal Univ Hosp, Expt Med Res Inst, N-0407 Oslo, Norway
[3] Stavanger Univ, Dept Elect & Comp Engn, N-4068 Stavanger, Norway
[4] Univ Oslo, UUH, Univ Div, N-0407 Oslo, Norway
[5] Ullevaal Univ Hosp, Div Surg, N-0407 Oslo, Norway
[6] Ullevaal Univ Hosp, Prehosp Div, N-0407 Oslo, Norway
关键词
out-of-hospital CPR; automated external defibrillators (AED); heart arrest; arrhythmia; electrocardiography; CPR artefact filtering;
D O I
10.1016/j.resuscitation.2005.04.009
中图分类号
R4 [临床医学];
学科分类号
1002 [临床医学]; 100602 [中西医结合临床];
摘要
There has recently been an increased attention focused on the importance of reducing time without blood flow from chest compressions (no flow time, NFT) during cardiopulmonary resuscitation (CPR). In this study we have analyzed and quantified the NFTs during external automatic defibrillation in 105 cardiac arrest patients. We found that for around half of the time (about 10 min), these patients were not perfused. We have proposed methods to reduce NFT in connection with analyses and shocks. The key factors were rhythm analysis during ongoing CPR, capacitor charging during analysis, 1 min of CPR immediately after a shock (with rhythm analysis during CPR at the end of the 1 min), and distinguishing between asystole and organized rhythm in analyses to skip pulse check if asystole. The potential reduction in NFT using these methods was calculated theoretically and we found a reduction in the total NFT of about 4.5 and 1 min. respectively, in the subgroups Of patients having at least one shock and patients having received no shocks. In the present Study, the median NFT ratio could theoretically be reduced from 51% to 34% or 49% to 39% depending on if the patient would have a shockable rhythm or not. By introducing the proposed methods into an AED, the NFT would be significantly reduced, hopefully increasing the survival. (c) 2005 Elsevier Ireland Ltd. All rights reserved.
引用
收藏
页码:95 / 101
页数:7
相关论文
共 23 条
[1]
Amann A, 2001, Curr Opin Crit Care, V7, P152, DOI 10.1097/00075198-200106000-00003
[2]
Automated external defibrillation versus manual defibrillation for prolonged ventricular fibrillation: Lethal delays of chest compressions before and after countershocks [J].
Berg, RA ;
Hilwig, RW ;
Kern, KB ;
Sanders, AB ;
Xavier, LC ;
Ewy, GA .
ANNALS OF EMERGENCY MEDICINE, 2003, 42 (04) :458-467
[3]
Adverse hemodynamic effects of interrupting chest compressions for rescue breathing during cardiopulmonary resuscitation for ventricular fibrillation cardiac arrest [J].
Berg, RA ;
Sanders, AB ;
Kern, KB ;
Hilwig, RW ;
Heidenreich, JW ;
Porter, ME ;
Ewy, GA .
CIRCULATION, 2001, 104 (20) :2465-2470
[4]
Influence of cardiopulmonary resuscitation prior to defibrillation in patients with out-of-hospital ventricular fibrillation [J].
Cobb, LA ;
Fahrenbruch, CE ;
Walsh, TR ;
Copass, MK ;
Olsufka, M ;
Breskin, M ;
Hallstrom, AP .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1999, 281 (13) :1182-1188
[5]
Effects of interrupting precordial compressions on the calculated probability of defibrillation success during out-of-hospital cardiac arrest [J].
Eftestol, T ;
Sunde, K ;
Steen, PA .
CIRCULATION, 2002, 105 (19) :2270-2273
[6]
Effects of cardiopulmonary resuscitation on predictors of ventricular fibrillation defibrillation success during out-of-hospital cardiac arrest [J].
Eftestol, T ;
Wik, L ;
Sunde, K ;
Steen, PA .
CIRCULATION, 2004, 110 (01) :10-15
[7]
Predicting outcome of defibrillation by spectral characterization and nonparametric classification of ventricular fibrillation in patients with out-of-hospital cardiac arrest [J].
Eftestol, T ;
Sunde, K ;
Aase, SO ;
Husoy, JH ;
Steen, PA .
CIRCULATION, 2000, 102 (13) :1523-1529
[8]
Feasibility of shock advice analysis during CPR through removal of CPR artefacts from the human ECG [J].
Eilevstjonn, J ;
Eftestol, T ;
Aase, SO ;
Myklebust, H ;
Husoy, JH ;
Steen, PA .
RESUSCITATION, 2004, 61 (02) :131-141
[9]
Cardiopulmonary resuscitation - some physiological considerations [J].
Frenneaux, M .
RESUSCITATION, 2003, 58 (03) :259-265
[10]
Fibrillation power, an alternative method of ECG spectral analysis for prediction of countershock success in a porcine model of ventricular fibrillation [J].
Hamprecht, FA ;
Achleitner, U ;
Krismer, AC ;
Lindner, KH ;
Wenzel, V ;
Strohmenger, HU ;
Thiel, W ;
van Gunsteren, WF ;
Amann, A .
RESUSCITATION, 2001, 50 (03) :287-296