A novel heart/trunk simulator for the study ofelectromagnetic interference with active implantable devices

被引:3
作者
Angeloni, A [1 ]
Barbaro, V [1 ]
Bartolini, P [1 ]
Calcagnini, G [1 ]
Censi, F [1 ]
机构
[1] Ist Super Sanita, Biomed Engn Lab, I-00161 Rome, Italy
关键词
electromagnetic compatibility; cardiac implantable devices; heart/trunk; simulator;
D O I
10.1007/BF02345317
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
This paper describes a portable heart simulator for the study of electromagnetic interference with active implantable devices. The simulator consists of plexiglas(TM) box divided into three chambers simulating the left atrium and the ventricles, plus a lateral compartment for the implantable device. The box is linked to a laptop computer by an analogue-to-digital convertor board, and the three chambers are monitored and driven by dedicated hardware and software interfaces. Synthetic endocardial atrial and ventricle signals for 13 cardiac rhythms are stored in the computer. They are applied to the cardiac chambers by AgCl plates. Sensing electrodes are in the form of AgCl needles inserted in saline. The simulator was able to demonstrate the behaviour of three pacemakers tested in the absence and presence of electromagnetic interference, generated by mobile phones (European GSM 900 and 1800 MHz) that emitted up to 2 W (1 W at 1800 MHz). Pacemakers can be programmed with sensitivity from 0.1 mV to 5 mV, pulse width from 0.1 ms to 1.5 ms and pulse amplitude from 0.5 V to 5 V. The structural separation in three cardiac chambers (plus the one for the device) allowed a fast analysis procedure for dual- and tri-chamber implantable devices.
引用
收藏
页码:550 / 555
页数:6
相关论文
共 11 条
[1]   DO EUROPEAN GSM MOBILE CELLULAR PHONES POSE A POTENTIAL RISK TO PACEMAKER PATIENTS [J].
BARBARO, V ;
BARTOLINI, P ;
DONATO, A ;
MILITELLO, C ;
ALTAMURA, G ;
AMMIRATI, F ;
SANTINI, M .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1995, 18 (06) :1218-1224
[2]   Electromagnetic interference of analog cellular telephones with pacemakers [J].
Barbaro, V ;
Bartolini, P ;
Donato, A ;
Militello, C .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1996, 19 (10) :1410-1418
[3]   Electromagnetic interference of digital and analog cellular telephones with implantable cardioverter defibrillators: In vitro and in vivo studies [J].
Barbaro, V ;
Bartolini, P ;
Bellocci, F ;
Caruso, F ;
Donato, A ;
Gabrielli, D ;
Militello, C ;
Montenero, AS ;
Zecchi, P .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1999, 22 (04) :626-634
[4]   Cellular phone interference testing of implantable cardiac defibrillators in vitro [J].
Bassen, HI ;
Moore, HJ ;
Ruggera, PS .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1998, 21 (09) :1709-1715
[5]  
CARILLO R, 1995, PACE, V18, P863
[6]   In-vivo testing of digital cellular telephones in patients with implantable cardioverter-defibrillators [J].
Chiladakis, JA ;
Davlouros, P ;
Agelopoulos, G ;
Manolis, AS .
EUROPEAN HEART JOURNAL, 2001, 22 (15) :1337-1342
[7]   Digital cellular telephone interaction with implantable cardioverter-defibrillators [J].
Fetter, JG ;
Ivans, V ;
Benditt, DG ;
Collins, J .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1998, 31 (03) :623-628
[8]  
Furman S, 1968, Ann Thorac Surg, V6, P90
[9]   Wireless phones and pacemaker interaction [J].
Hayes, DL .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1996, 19 (10) :1405-1406
[10]   Electromagnetic interference of pacemakers by mobile phones [J].
Irnich, W ;
Batz, L ;
Muller, R ;
Tobisch, R .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1996, 19 (10) :1431-1446