A dual-adsorbent preconcentrator for a portable indoor-VOC microsensor system

被引:107
作者
Lu, CJ
Zellers, ET [1 ]
机构
[1] Univ Michigan, Dept Environm Hlth Sci, Ann Arbor, MI 48109 USA
[2] Univ Michigan, Dept Chem, Ann Arbor, MI 48109 USA
关键词
D O I
10.1021/ac001524k
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The development and testing of a miniature dual-adsorbent preconcentrator for a microsensor-based analytical system designed to determine complex volatile organic chemical (VOC) mixtures encountered in indoor working environments at low part-per-billion levels is described. Candidate adsorbents were screened for thermal-desorption bandwidth and breakthrough volume against 20 volatile organic vapors and subsets thereof as a function of several relevant variables. A glass capillary (1.1 mm i.d.) packed with 3.4 mg of Carbopack X and 1.2 mg of Carboxen 1000 provides sufficient capacity for a l-L dry-air sample containing all 20 vapors at concentrations of 100 ppb as well as providing a composite half-height peak width of < 3 s at:a desorption temperature of 300 degreesC and a now rate of 4 mL/min. Required adsorbent masses increase to 7.0 and 1.5 mg, respectively, for the same mixture at component concentrations of 1 ppm. Vapor breakthrough volumes for the Carbopack X are unaffected by humidity from 0 to 100% RH, but those for the Carboxen 1000 are significantly reduced, requiring an additional 0.9 mg of the latter to avoid premature breakthrough at the 100 ppb level. Good peak shapes, efficient chromatographic separation of preconcentrated sample mixture components, and detection limits in the low-parts-per-billion range using an integrated surface-acoustic-wave (SAW) sensor are achieved. Preconcentrator design and operating parameters are considered in terms of the vapor bed-residence times and breakthrough volumes in the context of the modified Wheeler equation.
引用
收藏
页码:3449 / 3457
页数:9
相关论文
共 53 条
[1]  
ACKLEY MW, 1987, AM IND HYG ASSOC J, V48, P447
[2]   CONCENTRATIONS OF VOLATILE ORGANIC-COMPOUNDS IN INDOOR AIR - A REVIEW [J].
BROWN, SK ;
SIM, MR ;
ABRAMSON, MJ ;
GRAY, CN .
INDOOR AIR-INTERNATIONAL JOURNAL OF INDOOR AIR QUALITY AND CLIMATE, 1994, 4 (02) :123-134
[3]  
BURTON B, 1997, VOLATILE ORGANIC COM, P127
[4]   THERMAL-DESORPTION EFFICIENCIES FOR DIFFERENT ADSORBATE ADSORBENT SYSTEMS TYPICALLY USED IN AIR MONITORING PROGRAMS [J].
CAO, XL ;
HEWITT, CN .
CHEMOSPHERE, 1993, 27 (05) :695-705
[5]  
COHEN HJ, 1993, AM IND HYG ASSOC J, V54, P432
[6]  
FENG CY, 2000, J ENVIRON MONITOR, V2, P313
[7]   THE MICRO-TRAP - AN ALTERNATIVE TO CRYOFOCUSING IN CAPILLARY GAS-CHROMATOGRAPHY [J].
FRANK, W ;
FRANK, H .
CHROMATOGRAPHIA, 1990, 29 (11-12) :571-574
[8]  
FRYE GC, 1994, EL SOC M MIAM BEACH, P1046
[9]  
Frye-Mason G, 2000, MESA MG, P229
[10]   Dry purge for the removal of water from the solid sorbents used to sample volatile organic compounds from the atmospheric air [J].
Gawlowski, J ;
Gierczak, T ;
Pietruszyñska, Z ;
Gawrys, M ;
Niedzielski, J .
ANALYST, 2000, 125 (11) :2112-2117