A comparative tribological study of the wear of composite cushion cups in a physiological hip joint simulator

被引:30
作者
Bigsby, RJA
Auger, DD
Jin, ZM
Dowson, D
Hardaker, CS
Fisher, J [1 ]
机构
[1] Univ Leeds, Dept Mech Engn, Leeds L52 9JT, W Yorkshire, England
[2] DePuy Int Ltd, Leeds, W Yorkshire, England
基金
英国工程与自然科学研究理事会;
关键词
composite cushion bearing; wear; UHMWPE; hip joint simulator;
D O I
10.1016/S0021-9290(98)00034-7
中图分类号
Q6 [生物物理学];
学科分类号
071011 [生物物理学];
摘要
A composite cushion acetabular cup for a total hip replacement has been designed and developed jointly by Leeds University and DePuy International. In order to assess the long-term performance of this novel design, two sets of simulator tests of more than 4 million cycles duration have been carried out with the cushion bearings using the Leeds PA hip joint simulator with bovine serum as the lubricant. The results of these simulator tests were compared to the results from a previously reported study that used 32 mm ultrahigh molecular weight polyethylene (UHMWPE) acetabular cups. Under a physiological walking cycle simulation, with continuous cyclic motion and loading, the composite cushion cups produced negligible wear compared to a volumetric wear rate of 32 mm(3) per million cycles for the conventional UHMWPE acetabular cups. This study has demonstrated for the first time the beneficial effects of fluid film lubrication in reducing wear in composite cushion acetabular cups. (C) 1998 Published by Elsevier Science Ltd. All rights reserved.
引用
收藏
页码:363 / 369
页数:7
相关论文
共 14 条
[1]
AMSTUTZ HC, 1992, CLIN ORTHOP RELAT R, P7
[2]
Auger D D, 1993, Proc Inst Mech Eng H, V207, P25, DOI 10.1243/PIME_PROC_1993_207_265_02
[3]
BIGSBY RJA, 1996, J ENG MED, V211, P265
[4]
DERBYSHIRE B, 1994, P I MECH ENG H, V208, P151
[5]
Dowson D, 1991, Proc Inst Mech Eng H, V205, P59, DOI 10.1243/PIME_PROC_1991_205_269_02
[6]
AN EVALUATION OF THE PENETRATION OF CERAMIC FEMORAL HEADS INTO POLYETHYLENE ACETABULAR CUPS [J].
DOWSON, D ;
JOBBINS, B ;
SEYEDHARRAF, A .
WEAR, 1993, 162 (pt B) :880-889
[7]
DOWSON D, 1988, ENG MED, V17, P1
[8]
HARRIS WH, 1995, CLIN ORTHOP RELAT R, P46
[9]
Jin Z M, 1991, Proc Inst Mech Eng H, V205, P219, DOI 10.1243/PIME_PROC_1991_205_297_02
[10]
McClure G, 1996, Proc Inst Mech Eng H, V210, P89, DOI 10.1243/PIME_PROC_1996_210_397_02