Influence of static pressure and shear rate on hemolysis of red blood cells

被引:20
作者
Yasuda, T [1 ]
Funakubo, A
Miyawaki, F
Kawamura, T
Higami, T
Fukui, Y
机构
[1] Tokyo Denki Univ, Frontier R&D Ctr, Tokyo 101, Japan
[2] Tokyo Denki Univ, Dept Elect & Comp Engn, Tokyo 101, Japan
[3] Tokyo Denki Univ, Dept Biotechnol, Tokyo 101, Japan
[4] Hyogo Brain & Heart Ctr, Div Cardiovasc Surg, Himeji, Hyogo, Japan
关键词
All Open Access; Bronze;
D O I
10.1097/00002480-200107000-00011
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The purpose of this study was to investigate the effect of multiple mechanical forces in hemolysis. Specific attention is focused on the effects of sheer and pressure. An experimental apparatus consisting of a rotational viscometer, compression chamber, and heat exchanger was prepared to apply multiple mechanical forces to a blood sample. The rotational viscometer, in which bovine blood was subjected to sheer rates of 0, 500, 1,000, and 1,500 s(-1), was set in the compression chamber and pressurized with an air compressor at 0, 200, 400, and 500 mm Hg. The blood temperature was maintained at 21 degreesC and 28 degreesC. Free hemoglobin at 600 mm Hg was observed to be approximately four times higher than at 0 mm Hg for a shear rate of 1,500 s(-1) (p < 0.05). The results suggest that the increase in hemolysis is strongly related to pressure when high shear rates are applied to the erythrocytes. The data acquired in this study will be helpful in the development of artificial organs, where it will facilitate the prediction of hemolysis in flow dynamics analysis, flow visualization, and computational fluid dynamics.
引用
收藏
页码:351 / 353
页数:3
相关论文
共 9 条
[1]  
BLACKSHEAR PL, 1965, T AM SOC ART INT ORG, V11, P112
[2]   Effects of static pressure on red blood cells on removal of the air interface [J].
Chambers, SD ;
Laberteaux, KR ;
Merz, SI ;
Montoya, JP ;
Bartlett, RH .
ASAIO JOURNAL, 1996, 42 (06) :947-950
[3]  
Handa N., 1998, Japanese Journal of Artificial Organs, V27, P118
[4]  
HASHIMOTO S, 1985, BIOMATERIALS, V3, P81
[5]   RED BLOOD-CELL DAMAGE BY SHEAR-STRESS [J].
LEVERETT, LB ;
LYNCH, EC ;
ALFREY, CP ;
HELLUMS, JD .
BIOPHYSICAL JOURNAL, 1972, 12 (03) :257-&
[6]   Hemolytic effect of surface roughness of an impeller in a centrifugal blood pump [J].
Takami, Y ;
Nakazawa, T ;
Makinouchi, K ;
Tayama, E ;
Glueck, J ;
Benkowski, R ;
Nose, Y .
ARTIFICIAL ORGANS, 1997, 21 (07) :686-690
[7]  
YAMANE T, 1997, ASAIO J, V43, P635
[8]  
Yasuda T., 1996, Japanese Journal of Artificial Organs, V25, P632
[9]   A MATHEMATICAL-MODEL FOR SHEAR-INDUCED HEMOLYSIS [J].
YELESWARAPU, KK ;
ANTAKI, JF ;
KAMENEVA, MV ;
RAJAGOPAL, KR .
ARTIFICIAL ORGANS, 1995, 19 (07) :576-582