Tolerances for the accuracy of photon beam dose calculations of treatment planning systems

被引:197
作者
Venselaar, J
Welleweerd, H
Mijnheer, B
机构
[1] Dr B Vverbeeten Inst, Dept Radiotherapy, NL-5000 LA Tilburg, Netherlands
[2] Univ Med Ctr, Dept Radiotherapy, NL-3508 GA Utrecht, Netherlands
[3] Netherlands Canc Inst, Antoni Van Leeuwenhoek Hosp, Dept Radiotherapy, NL-1066 CX Amsterdam, Netherlands
关键词
quality assurances; treatment planning systems; photon beams; dose calculation;
D O I
10.1016/S0167-8140(01)00377-2
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background and purpose: To design a consistent set of criteria for acceptability of photon beam dose calculations of treatment planning systems. The set should be applicable in combination with a test package used for evaluation of a treatment planning system, such as the ones proposed by the AAPM Task Group 23 or by the Netherlands Commission on Radiation Dosimetry. Results: Tolerances have been defined for the accuracy with which a treatment planning system should be able to calculate the dose in different parts of a photon beam: the central beam axis and regions with large and small dose gradients. For increasing complexity of the geometry, wider tolerances are allowed, varying between 2 and 5%. For the evaluation of a large number of data points an additional quantity, the confidence limit, has been introduced, which combines the influence of systematic and random deviations. The proposed tolerances have been compared with other recommendations for a number of clinically relevant examples, showing considerable differences, which are partly due to the way the complexity of the geometry is taken into account. Furthermore differences occur if criteria for acceptability of dose calculations are related either to the local dose value or to a normalized dose value. Conclusions: Although it is acknowledged that the general aim must be to have good agreement between dose calculation and the actual dose value, e.g. within 2% or 2 mm, current day algorithms and their implementation into commercial treatment planning systems result often in larger deviations. A high accuracy can at present only be achieved in relatively simple cases. The new set of tolerances and the quantity confidence limit have proven to be useful tools for the acceptance of photon beam dose calculation algorithms of treatment planning systems. (C) 2001 Elsevier Science Ireland Ltd. All rights reserved.
引用
收藏
页码:191 / 201
页数:11
相关论文
共 45 条
[1]   Application of AAPM Radiation Therapy Committee Task Group 23 test package for comparison of two treatment planning systems for photon external beam radiotherapy [J].
Alam, R ;
Ibbott, GS ;
Pourang, R ;
Nath, R .
MEDICAL PHYSICS, 1997, 24 (12) :2043-2054
[2]  
*AM ASS PHYS MED, 1995, 55 AAPM TASK GROUP 2
[3]  
Brahme A., 1988, ACTA ONCOL S, V1, P1
[4]   A method to check the accuracy of dose computation using quality index: Application to scatter contribution in high energy photon beams [J].
Caneva, S ;
Rosenwald, JC ;
Zefkili, S .
MEDICAL PHYSICS, 2000, 27 (05) :1018-1024
[5]   Dosimetric comparison of treatment planning systems in irradiation of breast with tangential fields [J].
Cheng, CW ;
Das, IJ ;
Tang, W ;
Chang, S ;
Tsai, JS ;
Ceberg, C ;
DeGaspie, B ;
Singh, R ;
Fien, DA ;
Fowble, B .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1997, 38 (04) :835-842
[6]  
Coffey CW, 1985, MED DOS, V10, P9
[7]  
Cygler J, 1988, Med Dosim, V13, P155
[8]  
DAHLIN H, 1983, ACTA RADIOL ONCOL, V22, P398
[10]  
Damen EMF, 1997, PROCEEDINGS OF THE XIITH INTERNATIONAL CONFERENCE ON THE USE OF COMPUTERS IN RADIATION THERAPY, P365