Development and evaluation of a three-dimensional hyperthermia applicator with water-coated antennas (WACOA)

被引:7
作者
Nadobny, J
Wlodarczyk, W
Westhoff, L
Gellermann, J
Rau, B
Mönich, G
Wust, P
机构
[1] Humboldt Univ, Charite Med Ctr, Ctr Radiat Med, D-13353 Berlin, Germany
[2] Humboldt Univ, Charite Med Ctr, Dept Surg & Surg Oncol, D-13125 Berlin, Germany
[3] Tech Univ Berlin, Fac Elect Engn & Comp Sci, D-10587 Berlin, Germany
关键词
3-D hyperthermia applicator; annular phased array system; FDTD method; impedance matching; SAR pattern steering;
D O I
10.1118/1.1589611
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A novel twelve-channel three-dimensional (3-D) hyperthermia applicator has been developed and evaluated, which consists of twelve separate WAter COated Antenna (WACOA) modules. The modules are arranged in three transversal antenna rings (sub-arrays) and are placed into an acrylic applicator frame as cartridge-like elements in a staggered arrangement. The operating frequency is 100 MHz. For the design of the applicator, the finite-difference time-domain (FDTD) method was used. The applicator's dimensions allow its placement into the gantry of a magnetic resonance (MR) tomograph. The WACOA modules are designed as MR-compatible specially shaped metallic cylindrical dipole structures that are placed into hermetically closed water-filled cassettes. Due to the design of the dipole structures, only a conventional coaxial feed circuitry is needed, and no external impedance matching networks are necessary. Instead, fine on-line impedance matching is realized using adjustable tuning rods and matching rings, both elements being parts of the radiating antenna structure. Experimental and numerical evaluations demonstrate a good stability of impedance matching, a low inter-channel coupling of less than -20 dB, and a good ability of field pattern steering. (C) 2003 American Association of Physicists in Medicine.
引用
收藏
页码:2052 / 2064
页数:13
相关论文
共 42 条
[1]   Transient neuromotor phenotype in transgenic spastic mice expressing low levels of glycine receptor β-subunit:: an animal model of startle disease [J].
Becker, L ;
Hartenstein, B ;
Schenkel, J ;
Kuhse, J ;
Betz, H ;
Weiher, H .
EUROPEAN JOURNAL OF NEUROSCIENCE, 2000, 12 (01) :27-32
[2]  
BREZOVICH IA, 1989, HYPERTHERMIC ONCOLOG, P809
[3]   NUMERICAL-SIMULATION OF ANNULAR-PHASED ARRAYS OF DIPOLES FOR HYPERTHERMIA OF DEEP-SEATED TUMORS [J].
CHEN, JY ;
GANDHI, OP .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1992, 39 (03) :209-216
[4]   DESIGN OF A CLINICAL DEEP-BODY HYPERTHERMIA SYSTEM BASED ON THE COAXIAL TEM APPLICATOR [J].
DELEEUW, AAC ;
LAGENDIJK, JJW .
INTERNATIONAL JOURNAL OF HYPERTHERMIA, 1987, 3 (05) :413-421
[5]  
Gandhi OP, 1998, ART H ANT PROP LIB, P613
[6]   Neoadjuvant chemotherapy combined with regional hyperthermia (RHT) for locally advanced primary or recurrent high-risk adult soft-tissue sarcomas (STS) of adults: long-term results of a phase II study [J].
Issels, RD ;
Abdel-Rahman, S ;
Wendtner, CM ;
Falk, MH ;
Kurze, V ;
Sauer, H ;
Aydemir, U ;
Hiddemann, W .
EUROPEAN JOURNAL OF CANCER, 2001, 37 (13) :1599-1608
[7]  
Kunz K. S., 1993, FINITE DIFFERENCE TI
[8]   A SIMPLE FDTD MODEL FOR TRANSIENT EXCITATION OF ANTENNAS BY TRANSMISSION-LINES [J].
MALONEY, JG ;
SHLAGER, KL ;
SMITH, GS .
IEEE TRANSACTIONS ON ANTENNAS AND PROPAGATION, 1994, 42 (02) :289-292
[9]  
Montoya TP, 1999, MICROW OPT TECHN LET, V21, P105, DOI 10.1002/(SICI)1098-2760(19990420)21:2<105::AID-MOP8>3.0.CO
[10]  
2-O