Corrosion of spinal implants retrieved from patients with scoliosis

被引:33
作者
Akazawa, T
Minami, S
Takahashi, K
Koiani, T
Hanawa, T
Moriya, H
机构
[1] Chiba Univ, Grad Sch Med, Dept Orthoped Surg, Chuo Ku, Chiba 2608670, Japan
[2] Natl Inst Mat Sci, Biomat Ctr, Tsukuba, Ibaraki, Japan
关键词
corrosion; spinal implants; long-term implantation;
D O I
10.1007/s00776-004-0867-3
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Spinal implants retrieved from 11 patients with scoliosis were examined. All the implants were posterior instrumentation systems made of 316L stainless steel and composed of rods, hooks, and crosslink connectors. Corrosion was classified into grades 0 to 3 based on macroscopic findings of the rod surface at the junction of each hook or crosslink connector. Grade 0 was defined as no sign of corrosion. grade 1 as surface discoloration, grade 2 as superficial metal loss. and grade 3 as severe metal loss. The depths and characteristics of metal loss areas were examined. Spinal implants showed more corrosion after long-term implantation than after short-term implantation. Corrosion was seen on many of the rod junctions (66.2%) after long-term implantation. but there was no difference between the junction at the hook and those at the crosslink connector. It is thought that intergranular corrosion and fretting contributed to the corrosion of implants. The current study demonstrated that corrosion takes place at many of the rod junctions in long-term implantation. We recommend removal of the spinal implants after solid bony union.
引用
收藏
页码:200 / 205
页数:6
相关论文
共 14 条
[1]
*ASTM, 1997, ANN BOOK ASTM STAND, V2, P237
[2]
ASTM, 1999, ANN B ASTM STAND, V03, P1, DOI [10.1520/G0106-89R10.2, DOI 10.1520/G0106-89R10.2]
[3]
CORROSION OF THE HARRINGTONS INSTRUMENTATION AND BIOLOGICAL BEHAVIOR OF THE ROD - HUMAN SPINE SYSTEM [J].
AULISA, L ;
DIBENEDETTO, A ;
VINCIGUERRA, A ;
LORINI, G ;
TRANQUILLILEALI, P .
BIOMATERIALS, 1982, 3 (04) :246-+
[4]
COOK SD, 1985, CLIN ORTHOP RELAT R, P236
[5]
Metal sensitivity in patients with orthopaedic implants [J].
Hallab, N ;
Merritt, K ;
Jacobs, JJ .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2001, 83A (03) :428-436
[6]
XPS characterization of the surface oxide film 316L stainless steel samples that were located in quasi-biological environments [J].
Hanawa, T ;
Hiromoto, S ;
Yamamoto, A ;
Kuroda, D ;
Asami, K .
MATERIALS TRANSACTIONS, 2002, 43 (12) :3088-3092
[7]
Corrosion of metal orthopaedic implants [J].
Jacobs, JJ ;
Gilbert, JL ;
Urban, RM .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1998, 80A (02) :268-282
[8]
CORROSION OF MODULAR HIP PROSTHESES [J].
MATHIESEN, EB ;
LINDGREN, JU ;
BLOMGREN, GGA ;
REINHOLT, FP .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1991, 73 (04) :569-575
[9]
EFFECT OF PROTEINS AND PH ON FRETTING CORROSION AND METAL-ION RELEASE [J].
MERRITT, K ;
BROWN, SA .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1988, 22 (02) :111-120
[10]
CHIBA SPINAL SYSTEM IN THE OPERATIVE MANAGEMENT OF SCOLIOSIS [J].
NAKATA, Y ;
MORIYA, H ;
KITAHARA, H ;
MINAMI, S ;
TAKAHASHI, K ;
OHTSUKA, Y .
SPINE, 1992, 17 (10) :1166-1173