Microtrap modulated flame ionization detector for on-line monitoring of methane

被引:20
作者
Thammakhet, C
Thavarungkul, P
Brukh, R
Mitra, S
Kanatharana, P [1 ]
机构
[1] Prince Songkla Univ, Dept Chem, Fac Sci, Analyt & Environm Trace Anal Res Unit, Hat Yai, Songkhla, Thailand
[2] Prince Songkla Univ, Dept Phys, Fac Sci, Biophys Res Unit Biosensors & Biocurrents, Hat Yai, Songkhla, Thailand
[3] New Jersey Inst Technol, Dept Chem & Environm Sci, Newark, NJ 07102 USA
关键词
methane; microtrap; carbosphere; high purity gases;
D O I
10.1016/j.chroma.2005.03.041
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A microtrap for on-line analysis of trace methane was developed. Silicosteel tubing, 15 cm long, 1.02 turn i.d. and 1.59 mm o.d. was packed with Carbosphere 80/100 mesh and placed between the standard gas source and the flame ionization detector (FID) to be used as a preconcentrator and an injector. The sample stream was passed continuously through the microtrap that was heated at a fixed interval by an electric pulse. To achieve the highest adsorption and desorption efficiency, parameters affecting the response of the system were optimized, and adsorption temperature in the range of 25 to -50 degrees C was studied. The response from the microtrap as compared to that from direct-flame ionization detector was enhanced by as much as 260 times. The on-line microtrap system showed good linearity (r(2) > 0.99), low detection limit (28.3 ppbv) and good long term stability (relative standard deviation, RSD of less than 5.0 %). Therefore, this simple device is suitable for on-line analysis of trace methane and similar small molecules in the environment as in high purity gases. (c) 2005 Elsevier B.V. All rights reserved.
引用
收藏
页码:243 / 248
页数:6
相关论文
共 16 条
[11]   CONTINUOUS GAS-CHROMATOGRAPHIC MONITORING OF LOW CONCENTRATION SAMPLE STREAMS USING AN ONLINE MICROTRAP [J].
MITRA, S ;
YUN, C .
JOURNAL OF CHROMATOGRAPHY, 1993, 648 (02) :415-421
[12]   Continuous monitoring of volatile organic compounds in air emissions using an on-line membrane extraction-microtrap-gas chromatographic system [J].
Mitra, S ;
Zhu, NH ;
Zhang, X ;
Kebbekus, B .
JOURNAL OF CHROMATOGRAPHY A, 1996, 736 (1-2) :165-173
[13]   REPETITIVE LIQUID INJECTION SYSTEM FOR INVERSE GAS-CHROMATOGRAPHY [J].
PANDA, S ;
BU, Q ;
YUN, KS ;
PARCHER, JF .
JOURNAL OF CHROMATOGRAPHY A, 1995, 715 (02) :279-285
[14]  
Rafson HJ., 1998, ODOR VOC CONTROL HDB
[15]   BREAKTHROUGH VOLUME OF MONOTERPENES ON TENAX TA - INFLUENCE OF TEMPERATURE AND CONCENTRATION FOR ALPHA-PINENE [J].
SIMON, V ;
RIBA, ML ;
WALDHART, A ;
TORRES, L .
JOURNAL OF CHROMATOGRAPHY A, 1995, 704 (02) :465-471
[16]  
2003, TIGEROPTICS TRACE ME