Influence of OSEM and segmented attenuation correction in the calculation of standardised uptake values for [18F]FDG PET

被引:70
作者
Visvikis, D
Cheze-LeRest, C
Costa, DC
Bomanji, J
Gacinovic, S
Ell, PJ
机构
[1] UCL Royal Free & Univ Coll Med Sch, Middlesex Hosp, Inst Nucl Med, London W1T 3AA, England
[2] Univ Bretagne Occidentale, Fac Med, LATIM, Biophys Lab, F-29285 Brest, France
关键词
ordered subsets; expectation maximisation; segmented attenuation correction; standardised uptake values;
D O I
10.1007/s002590100566
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Standardised Uptake Values (SUVs) are widely used in positron emission tomography (PET) as a semi-quantitative index of fluorine-18 labelled fluorodeoxyglucose uptake. The objective of this study was to investigate any bias introduced in the calculation of SUVs as a result of employing ordered subsets-expectation maximisation (OSEM) image reconstruction and segmented attenuation correction (SAC). Variable emission and transmission time durations were investigated. Both a phantom and a clinical evaluation of the bias were carried out. The software implemented in the GE Advance PET scanner was used. Phantom studies simulating tumour imaging conditions were performed. Since a variable count rate may influence the results obtained using OSEM, similar acquisitions were performed at total count rates of 34 kcps and 12 kcps. Clinical data consisted of 100 patient studies. Emission datasets of 5 and 15 min duration were combined with 15-, 3-, 2- and 1-min transmission datasets for the reconstruction of both phantom and patient studies. Two SUVs were estimated using the average (SUVavg) and the maximum (SUVmax) count density from regions of interest placed well inside structures of interest. The percentage bias of these SUVs compared with the values obtained using a reference image was calculated. The reference image was considered to be the one produced by filtered backprojection (FBP) image reconstruction with measured attenuation correction using the 15-min emission and transmission datasets for each phantom and patient study. A bias of 5%-20% was found for the SUVavg and SUVmax in the case of FBP with SAC using variable transmission times. In the case of OSEM with SAC, the bias increased to 10%-30%. An overall increase of 5%-10% was observed with the use of SUVmax. The 5-min emission dataset led to an increase in the bias of 25%-100%. with the larger increase recorded for the SUVmax. The results suggest that OSEM and SAC with 3 and 2 min transmission may be reliably used to reduce the overall data acquisition time without compromising the accuracy of SUVs.
引用
收藏
页码:1326 / 1335
页数:10
相关论文
共 19 条
  • [1] A LEAST BIASED FUZZY CLUSTERING METHOD
    BENI, G
    LIU, XM
    [J]. IEEE TRANSACTIONS ON PATTERN ANALYSIS AND MACHINE INTELLIGENCE, 1994, 16 (09) : 954 - 960
  • [2] An automatic classification technique for attenuation correction in positron emission tomography
    Bettinardi, V
    Pagani, E
    Gilardi, MC
    Landoni, C
    Riddell, C
    Rizzo, G
    Castiglioni, I
    Belluzzo, D
    Lucignani, G
    Schubert, S
    Fazio, F
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE, 1999, 26 (05) : 447 - 458
  • [3] Clinical role of positron emission tomography in oncology
    Bomanji, J. B.
    Costa, D. C.
    Ell, P. J.
    [J]. LANCET ONCOLOGY, 2001, 2 (03) : 157 - 164
  • [4] POST-INJECTION TRANSMISSION ATTENUATION MEASUREMENTS FOR PET
    DAUBEWITHERSPOON, ME
    CARSON, RE
    GREEN, MV
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1988, 35 (01) : 757 - 761
  • [5] Eary JF, 1998, J NUCL MED, V39, P250
  • [6] ACCELERATED IMAGE-RECONSTRUCTION USING ORDERED SUBSETS OF PROJECTION DATA
    HUDSON, HM
    LARKIN, RS
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 1994, 13 (04) : 601 - 609
  • [7] JORDAN K, EEC EMISSION PHANTOM
  • [8] SINGLES TRANSMISSION IN VOLUME-IMAGING PET WITH A CS-137 SOURCE
    KARP, JS
    MUEHLLEHNER, G
    QU, H
    YAN, XH
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1995, 40 (05) : 929 - 944
  • [9] A clinical evaluation of the quantitative accuracy of simultaneous emission/transmission scanning in whole-body positron emission tomography
    Lodge, MA
    Badawi, RD
    Marsden, PK
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE, 1998, 25 (04) : 417 - 423
  • [10] A PET study of 18FDG uptake in soft tissue masses
    Lodge, MA
    Lucas, JD
    Marsden, PK
    Cronin, BF
    O'Doherty, MJ
    Smith, MA
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE, 1999, 26 (01) : 22 - 30