Measurements of Gamma-Knife helmet output factors using a radiophotoluminescent glass rod dosimeter and a diode detector

被引:77
作者
Araki, F [1 ]
Ikegami, T
Ishidoya, T
Kubo, HD
机构
[1] Kumamoto Univ, Coll Med Sci, Dept Radiol Technol, Kumamoto 8620976, Japan
[2] Asahi Techno Glass Corp, Shizuoka 4210302, Japan
[3] Univ Calif Davis, Ctr Canc, Dept Radiat Oncol, Sacramento, CA 95817 USA
关键词
radiophotoluminescence; glass rod dosimeter; p-type silicon diode detector; Gamma-Knife; helmet output factors; angular dependence;
D O I
10.1118/1.1587451
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A radiophotoluminescent (RPL) glass rod dosimeter (GRD) and a small active volume p-type silicon diode detector are used for the measurement of the output factors from Gamma-Knife fields. The GRD system consists of small rod-shaped glass chip detectors and an automatic readout device. The output factors measured with the GRD from the 14, 8 and 4 mm helmets relative to the 18 mm helmet are 0.981, 0.942 and 0.877, respectively. Similarly, the corresponding output factors measured with the p-type silicon diode detector are 0.980, 0.949 and 0.867, respectively. The output factors are corrected for the end effect for each helmet. The output factors obtained from both detectors are in good agreement with the values in a recent publication and the values recommended by Elekta, the manufacturer. The directional dependence of these detectors is also measured. For the Gamma-Knife angle ranging from 6 to 36 degrees in the y-z plane of the stereotactic space, the measured angular dependence of the GRD is approximately 1.0% at a 4 MV x-ray beam. The response of the silicon diode detector indicates approximately 3-4% directional dependence for the same angular range for a 6 MV x-ray beam. The Gamma-Knife helmet output factors measured with the silicon diode detector are corrected for angular dependence. (C) 2003 American Association of Physicists in Medicine.
引用
收藏
页码:1976 / 1981
页数:6
相关论文
共 9 条
[1]  
*AS TECHN GLASS CO, 2000, EXPL MAT RPL GLASS D
[2]   A film technique for the determination of output factors and end effect times for the Leksell Gamma Knife [J].
Ekstrand, KE ;
Bourland, JD .
PHYSICS IN MEDICINE AND BIOLOGY, 2001, 46 (03) :703-706
[3]  
*ELEKT INSTR AB, 1998, LEKS GAMM KNIF B
[4]   Shuttle dose at the Vienna Leksell Gamma Knife [J].
Ertl, A ;
Zehetmayer, M ;
Schoggl, A ;
Kitz, K ;
Koschuch, R ;
Stadtmann, H ;
Kindl, P .
PHYSICS IN MEDICINE AND BIOLOGY, 1998, 43 (06) :1567-1578
[5]   Application of a newly developed photoluminescence glass dosimeter for measuring the absorbed dose in individual mice exposed to low-dose rate 137Cs γ-rays [J].
Hoshi, Y ;
Nomura, T ;
Oda, T ;
Iwasaki, T ;
Fujita, K ;
Ishikawa, T ;
Kato, A ;
Ikegami, T ;
Sakai, K ;
Tanooka, H ;
Yamada, T .
JOURNAL OF RADIATION RESEARCH, 2000, 41 (02) :129-137
[6]   Dosimetry and mechanical accuracy of the first rotating gamma system installed in North America [J].
Kubo, HD ;
Araki, F .
MEDICAL PHYSICS, 2002, 29 (11) :2497-2505
[7]   Precision dosimetry for narrow photon beams used in radiosurgery -: Determination of Gamma Knife® output factors [J].
Mack, A ;
Scheib, SG ;
Major, J ;
Gianolini, S ;
Pazmandi, G ;
Feist, H ;
Czempiel, H ;
Kreiner, HJ .
MEDICAL PHYSICS, 2002, 29 (09) :2080-2089
[8]  
*SCAND MED AB, 2000, EXPL MAT HI PSI FIEL
[9]  
Tsuda Masayuki, 2000, Igaku Butsuri, V20, P131