Candidate Reference Measurement Procedures for Serum 25-Hydroxyvitamin D3 and 25-Hydroxyvitamin D2 by Using Isotope-Dilution Liquid Chromatography-Tandem Mass Spectrometry

被引:188
作者
Stepman, Hedwig C. M. [1 ]
Vanderroost, An
Van Uytfanghe, Katleen [1 ]
Thienpont, Linda M. [1 ]
机构
[1] Univ Ghent, Analyt Chem Lab, Fac Pharmaceut Sci, B-9000 Ghent, Belgium
关键词
VITAMIN-D STATUS; CIRCULATING; 25-HYDROXYVITAMIN-D; ACCURATE; ASSAYS;
D O I
10.1373/clinchem.2010.152553
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
BACKGROUND: 25-hydroxyvitaminD[25(OH)D] assays are characterized by poor between-assay comparability. This result emphasizes the need for reference measurement procedures (RMPs) to establish calibration traceability and assist in method validation. We aimed at developing candidate RMPs on the basis of isotope-dilution liquid chromatography-tandem mass spectrometry (ID-LC-MS/MS) for separate quantification of serum 25(OH)D-2 and 25(OH)D-3. METHODS: Hexa-deuterated 25(OH)D-3/D-2 was added to serum. This mixture was extracted with n-hexane and fractionated on Sephadex LH-20 before 2-dimensional LC-MS/MS. In the first dimension, both procedures used a C4 column; however, in the second dimension, the 25(OH)D-2 procedure used a C18 and the 25(OH)D-3 procedure used a Zorbax SB-CN column. Calibration was traceable to the NIST Standard Reference Material (SRM) 2972. Validation comprised assessment of interference and limit of quantification/detection. Imprecision and trueness were validated by analysis of the SRM 972 against specifications (CV < 5% and bias < 1.7%). The expanded uncertainty for quadruplicate measurements was estimated. RESULTS: Testing of potentially interfering substances was negative. Interference by 3-epi-25(OH)D-3 was resolved by sufficient chromatographic resolution. The limits of quantification/detection were 1.1 nmol/L and 0.09 pmol/L for 25(OH)D-3 and 1.2 nmol/L and 0.05 pmol/L for 25(OH)D-2. Mean total CVs and differences from the SRM 972 target (+/- 1-sided 95% CI) were 2.1% and 1.1% +/- 1.5% [25(OH)D-3] and 3% and 1.3% +/- 0.6% [25(OH)D-2], respectively. The respective expanded uncertainties were 3.4% and 3.9%. CONCLUSIONS: From the validation data, we conclude that we achieved our objective of 2 state-of-the-art candidate RMPs for serum 25(OH)D-3 and 25(OH)D-2. (C) 2010 American Association for Clinical Chemistry
引用
收藏
页码:441 / 448
页数:8
相关论文
共 36 条
[31]  
2-X
[32]  
U.S. Department of Health and Human Services. Agency for Healthcare Research and Quality, VIT D CALC SYST REV
[33]   Traceability in Laboratory Medicine [J].
Vesper, Hubert W. ;
Thienpont, Linda M. .
CLINICAL CHEMISTRY, 2009, 55 (06) :1067-1075
[34]   Candidate reference method for the quantification of circulating 25-hydroxyvitamin D3 by liquid chromatography-tandem mass spectrometry [J].
Vogeser, M ;
Kyriatsoulis, A ;
Huber, E ;
Kobold, U .
CLINICAL CHEMISTRY, 2004, 50 (08) :1415-1417
[35]   Measurement of 25-hydroxyvitamin D in the clinical laboratory: Current procedures, performance characteristics and limitations [J].
Wallace, A. M. ;
Gibson, S. ;
de la Hunty, A. ;
Lamberg-Allardt, C. ;
Ashwell, M. .
STEROIDS, 2010, 75 (07) :477-488
[36]  
Yates AM, 2008, CLIN CHEM, V54, pA3