Development of a flow-through immunoassay system

被引:35
作者
Abdel-Hamid, I
Atanasov, P
Ghindilis, AL
Wilkins, E [1 ]
机构
[1] Univ New Mexico, Dept Chem & Nucl Engn, Farris Engn Ctr 209, Albuquerque, NM 87131 USA
[2] Synect Technol Corp, Denville, NJ 07834 USA
关键词
amperometric immunosensor; immunoassay system; flow-through immunoassay;
D O I
10.1016/S0925-4005(98)00125-7
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A semi-automated how-through immunoassay system consisting of an amperometric immunosensor and reagent flow arrangements has been developed. The amperometric immunosensor is based on a high-surface-area carbon immunoelectrode. The novel principle of flow immunoelectrode based on highly dispersed carbon material which acts as both an immunosorbent and an immunoelectrode was employed. The immunoelectrode is designed to be a disposable sensing element. The 'sandwich' scheme of immunoassay has been used and iodine formed as a result of the enzymatic oxidation of iodide by peroxidase-label has been detected amperometrically. The overall time of analysis including flowing of analyte, flowing of antigen, washing and detecting stages is as low as 17 min. The developed system allows fast determination of rabbit IgG (used as a model analyte) with a low detection limit in the 10(-11) M range and may easily be adapted for detection of various other analyses. The immunoassay system has the potential for miniaturization and complete automation allowing its use in laboratory setting as well as in field conditions. (C) 1998 Elsevier Science S.A. All rights reserved.
引用
收藏
页码:202 / 210
页数:9
相关论文
共 38 条
  • [1] BIOELECTROCATALYSIS BY IMMOBILIZED PEROXIDASE - THE REACTION-MECHANISM AND THE POSSIBILITY OF ELECTROANALYTICAL DETECTION OF BOTH INHIBITORS AND ACTIVATORS OF ENZYME
    BOGDANOVSKAYA, VA
    FRIDMAN, VA
    TARASEVICH, MR
    SCHELLER, F
    [J]. ANALYTICAL LETTERS, 1994, 27 (15) : 2823 - 2847
  • [2] BUCKLEY E, 1989, Analytical Proceedings, V26, P5
  • [3] Electrochemical immunosensors for determination of the pesticides 2,4-dichlorophenoxyacetic and 2,4,5-trichlorophenoxyacetic acids
    Dzantiev, BB
    Zherdev, AV
    Yulaev, MF
    Sitdikov, RA
    Dmitrieva, NM
    Moreva, IY
    [J]. BIOSENSORS & BIOELECTRONICS, 1996, 11 (1-2) : 179 - 185
  • [4] MEASUREMENT OF ALKALINE-PHOSPHATASE ACTIVITY BY ELECTROCHEMICAL DETECTION OF PHOSPHATE-ESTERS - APPLICATION TO AMPEROMETRIC ENZYME-IMMUNOASSAY
    FREW, JE
    FOULDS, NC
    WILSHERE, JM
    FORROW, NJ
    GREEN, MJ
    [J]. JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1989, 266 (02): : 309 - 316
  • [5] Flow-through amperometric immunosensor: Fast 'sandwich' scheme immunoassay
    Ghindilis, AL
    Krishnan, R
    Atanasov, P
    Wilkins, E
    [J]. BIOSENSORS & BIOELECTRONICS, 1997, 12 (05) : 415 - 423
  • [6] Potentiometric immunoelectrode for fast assay based on direct electron transfer catalyzed by peroxidase
    Ghindilis, AL
    Atanasov, P
    Wilkins, E
    [J]. SENSORS AND ACTUATORS B-CHEMICAL, 1996, 34 (1-3) : 528 - 532
  • [7] DETERMINATION OF P-AMINOPHENOL AND CATECHOLAMINES AT PICOMOLAR CONCENTRATIONS BASED ON RECYCLING ENZYME AMPLIFICATION
    GHINDILIS, AL
    MAKOWER, A
    BAUER, CG
    BIER, FF
    SCHELLER, FW
    [J]. ANALYTICA CHIMICA ACTA, 1995, 304 (01) : 25 - 31
  • [8] GHINDILIS AL, 1995, ANAL LETT, V28, P1
  • [9] NONCOMPETITIVE FLOW-INJECTION IMMUNOASSAY FOR A HAPTEN, ALPHA-(DIFLUOROMETHYL)ORNITHINE
    GUNARATNA, PC
    WILSON, GS
    [J]. ANALYTICAL CHEMISTRY, 1993, 65 (09) : 1152 - 1157
  • [10] SIMULTANEOUS IMMUNOASSAY USING ELECTROCHEMICAL DETECTION OF METAL-ION LABELS
    HAYES, FJ
    HALSALL, HB
    HEINEMAN, WR
    [J]. ANALYTICAL CHEMISTRY, 1994, 66 (11) : 1860 - 1865