Upconverting nanoparticle based optical sensor for carbon dioxide

被引:113
作者
Ali, Reham [1 ]
Saleh, Sayed M. [1 ]
Meier, Robert J. [1 ]
Azab, Hassan A. [2 ]
Abdelgawad, Ibraheim I. [3 ]
Wolfbeis, Otto S. [1 ]
机构
[1] Univ Regensburg, Inst Analyt Chem Chemo & Biosensors, D-93040 Regensburg, Germany
[2] Suez Canal Univ, Fac Sci, Dept Chem, Ismailia, Egypt
[3] Suez Canal Univ, Fac Sci, Dept Chem, Suez, Egypt
来源
SENSORS AND ACTUATORS B-CHEMICAL | 2010年 / 150卷 / 01期
关键词
Optical sensor; Carbon dioxide; Nanoparticle; Upconversion; Luminescence; UP-CONVERSION; ALPHA-NAPHTHOLPHTHALEIN; COLORIMETRIC CHANGE; FLUORESCENT PORPHYRIN; CHEMICAL SENSORS; CO2; SENSOR; GAS SENSOR; PH; OXYGEN; FLUOROSENSOR;
D O I
10.1016/j.snb.2010.07.031
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
We demonstrate a novel optical sensor for carbon dioxide in concern rations between 0 and 3% The sensing scheme is based on the optical interrogation of a 12-mu m polystyrene (PS) film containing upconverting nanoparticles (UCNPs, 40-100 nm in size) of the type NaYF4 Yb,Er, and the longwave absorption pH probe bromothymol blue (BTB) in its anionic (blue) form PS is chosen as a matrix because it displays permeation selectivity for CO2 and rejects morons The color of BTB in the PS matrix depends on the partial pressure of CO2 gas the UCNPs are photoexcited with a 980-nm laser diode to give a green (542 nm) and a red (657 nm) emission whose intensity is screened off (depending on whether BTB is present in as blue 01 yellow form) due to an inner filter effect The luminescence intensities of the UCNPs at 542 nm and 657 nm increase with increasing concentration of CO2 The pH probe BTB (a sulfonate) is used in the form of a lipophilic ion pair with the tetrabutylammonium cation (TBA) The strong base tetraoctylammonium hydroxide is added to the system and acts as a base to convert BTB in its phenoxide (blue) form. but also creates a buffer system This is the first optical sensor for CO2 that is based on the use of UCNPs Its response time is similar to 10s on switching from pure argon gas to 1% CO2 in argon, the recovery time of the sensing film is similar to 180 s, and the detection limit is 0 11% of CO2 (C) 2010 Elsevier B V All rights reserved
引用
收藏
页码:126 / 131
页数:6
相关论文
共 51 条
[1]   Visible light emission upon near-infrared excitation in a transparent solution of nanocrystalline β-NaGdF4:Yb3+, Er3+ [J].
Aebischer, A ;
Heer, S ;
Biner, D ;
Krämer, K ;
Haase, M ;
Güdel, HU .
CHEMICAL PHYSICS LETTERS, 2005, 407 (1-3) :124-128
[2]   Optical CO2 sensor of the combination of colorimetric change of α-naphtholphthalein in poly(isobutyl methacrylate) and fluorescent porphyrin in polystyrene [J].
Amao, Y ;
Komori, T .
TALANTA, 2005, 66 (04) :976-981
[3]   An optical sensor with the combination of colorimetric change of α-naphtholphthalein and internal reference luminescent dye CO2 in water [J].
Amao, Y ;
Nakamura, N .
SENSORS AND ACTUATORS B-CHEMICAL, 2005, 107 (02) :861-865
[4]   Optical CO2 sensor with the combination of colorimetric change of α-naphtholphthalein and internal reference fluorescent porphyrin dye [J].
Amao, Y ;
Nakamura, N .
SENSORS AND ACTUATORS B-CHEMICAL, 2004, 100 (03) :347-351
[5]   Upconversion and anti-stokes processes with f and d ions in solids [J].
Auzel, F .
CHEMICAL REVIEWS, 2004, 104 (01) :139-173
[6]   Composite material for simultaneous and contactless luminescent sensing and imaging of oxygen and carbon dioxide [J].
Borisov, Sergey M. ;
Krause, Christian ;
Arain, Sarina ;
Wolfbeis, Otto S. .
ADVANCED MATERIALS, 2006, 18 (12) :1511-+
[7]   Surface Modification of Upconverting NaYF4 Nanoparticles with PEG-Phosphate Ligands for NIR (800 nm) Biolabeling within the Biological Window [J].
Boyer, John-Christopher ;
Manseau, Marie-Pascale ;
Murray, Jill I. ;
van Veggel, Frank C. J. M. .
LANGMUIR, 2010, 26 (02) :1157-1164
[8]   Highly sensitive and linear optical fiber carbon dioxide sensor based on sol-gel matrix doped with silica particles and HPTS [J].
Chu, Chen-Shane ;
Lo, Yu-Lung .
SENSORS AND ACTUATORS B-CHEMICAL, 2009, 143 (01) :205-210
[9]   Optical pH, oxygen and carbon dioxide monitoring using a microdialysis approach [J].
Cooney, CG ;
Towe, BC ;
Eyster, CR .
SENSORS AND ACTUATORS B-CHEMICAL, 2000, 69 (1-2) :183-188
[10]   Up-converting phosphor technology-based lateral flow assay for detection of Schistosoma circulating anodic antigen in serum [J].
Corstjens, Paul L. A. M. ;
van Lieshout, Lisette ;
Zuiderwijk, Michel ;
Kornelis, Dieuwke ;
Tanke, Hans J. ;
Deelder, Andre M. ;
van Dam, Govert J. .
JOURNAL OF CLINICAL MICROBIOLOGY, 2008, 46 (01) :171-176