Evaluation of a disposable prism for applanation tonometry

被引:16
作者
Desai, SP
Sivakumar, S
Fryers, PT
Pandurangrao, S
机构
[1] Doncaster Royal Infirm, Dept Ophthalmol, Doncaster DN2 5LT, England
[2] Doncaster Hlth Author, Directorate Hlth Policy & Publ Hlth, Doncaster, England
关键词
cross-infection; disinfection; intraocular pressure; tonometry;
D O I
10.1038/eye.2001.94
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Background Recently the Medical Devices Agency recommended that 'ophthalmic devices that touch the surface of the eye should be restricted to single use'. Aim To evaluate one such device: a disposable tonometer prism for routine applanation tonometry. Methods The intraocular pressure (IOP) of 100 consecutive patients from a general eye clinic (197 eyes) was measured with both a disposable and the standard Goldmann tonometer (Goldmann). The level of agreement between the two methods of clinical measurement was assessed and the sensitivity and specificity of the disposable prism in detecting clinically significant raised IOP estimated. Results The mean difference in IOP measured by the two different prisms was 0.44 mmHg with a standard deviation (SD) of 1.54. The mean IOP for the disposable prism was 19.51 mmHg (SD 6.53 mmHg). The mean IOP for the standard Goldmann tonometer prism was 19.07 mmHg (SD 6.64 mmHg). The sensitivity to detect IOP > 21 mmHg was 95.9% (95% confidence interval (CI): 86.0-99.5%) and the specificity of 93.9% (95% CI: 88.8-97.2%). It gave a positive predictive value of 83.9% (95% CI: 71.7-92.4%). Conclusion There was close agreement between the IOP measurements obtained by the disposable tonometer prism and the Goldmann device for high and low pressures. If replicated, the high sensitivity and specificity would justify its use in screening.
引用
收藏
页码:279 / 282
页数:4
相关论文
共 11 条
[1]   DETERIORATION IN THE ACCURACY OF THE PULSAIR NONCONTACT TONOMETER WITH USE - NEED FOR REGULAR CALIBRATION [J].
ATKINSON, PL ;
WISHART, PK ;
JAMES, JN ;
VERNON, SA ;
REID, F .
EYE, 1992, 6 :530-534
[2]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[3]   On the precision and reliability of IOP measurements [J].
Gilchrist, JM .
BRITISH JOURNAL OF OPHTHALMOLOGY, 1996, 80 (07) :586-587
[4]   REDUCING BIAS DURING INTRAOCULAR-PRESSURE MEASUREMENT [J].
GILLOW, JT ;
AGGARWAL, R .
BRITISH JOURNAL OF OPHTHALMOLOGY, 1995, 79 (11) :1057-1058
[5]  
KOO D, 1989, INFECT CONT HOSP EP, V10, P547
[6]   Creutzfeldt-Jakob disease and the eye. II. Ophthalmic and neuro-ophthalmic features [J].
Lueck, CJ ;
McIlwaine, GG ;
Zeidler, M .
EYE, 2000, 14 (3) :291-301
[7]  
Maldonado MJ, 1996, OPHTHALMOLOGY, V103, P815
[8]  
Smith CA, 1999, AM J OPHTHALMOL, V127, P77
[9]  
SUDESH S, 1993, ACTA OPHTHALMOL, V71, P185
[10]   A LARGE OUTBREAK OF EPIDEMIC KERATOCONJUNCTIVITIS - PROBLEMS IN CONTROLLING NOSOCOMIAL SPREAD [J].
WARREN, D ;
NELSON, KE ;
FARRAR, JA ;
HURWITZ, E ;
HIERHOLZER, J ;
FORD, E ;
ANDERSON, LJ .
JOURNAL OF INFECTIOUS DISEASES, 1989, 160 (06) :938-943