Flow-switching device for comprehensive two-dimensional gas chromatography

被引:84
作者
Bueno, PA [1 ]
Seeley, JV [1 ]
机构
[1] Oakland Univ, Dept Chem, Rochester, MI 48309 USA
关键词
gas chromatography; comprehensive two-dimensional; flow-switching instrumentation; diesel fuel; volatile organic compounds;
D O I
10.1016/j.chroma.2003.10.033
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A simple flow-switching device has been developed as a differential flow modulator for comprehensive two-dimensional gas chromatography (GC x GC). The device is assembled from tubing, four tee unions, and a solenoid valve. The solenoid valve is located outside the oven of the gas chromatograph and is not in the sample path. The modulation technique has no inherent temperature restrictions and passes 100% of the primary column effluent to the secondary column(s). Secondary peaks are produced with widths at half maximum less than 100 ms when operating in GC x 2GC mode with a 2.0 s modulation period. The efficacy of this approach is demonstrated through the analysis of a standard mixture of volatile organic compounds (VOCs) and diesel fuel. (C) 2003 Elsevier B.V. All rights reserved.
引用
收藏
页码:3 / 10
页数:8
相关论文
共 17 条
[1]   Single-jet, single-stage cryogenic modulator for comprehensive two-dimensional gas chromatography (GC x GC) [J].
Adahchour, M ;
Beens, J ;
Brinkman, UAT .
ANALYST, 2003, 128 (03) :213-216
[2]   Simple, non-moving modulation interface for comprehensive two-dimensional gas chromatography [J].
Beens, J ;
Adahchour, M ;
Vreuls, RJJ ;
van Altena, K ;
Brinkman, UAT .
JOURNAL OF CHROMATOGRAPHY A, 2001, 919 (01) :127-132
[3]   Comprehensive two dimensional high-speed gas chromatography with chemometric analysis [J].
Bruckner, CA ;
Prazen, BJ ;
Synovec, RE .
ANALYTICAL CHEMISTRY, 1998, 70 (14) :2796-2804
[4]   AUTOMATED INSTRUMENTATION FOR COMPREHENSIVE 2-DIMENSIONAL HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY OF PROTEINS [J].
BUSHEY, MM ;
JORGENSON, JW .
ANALYTICAL CHEMISTRY, 1990, 62 (02) :161-167
[5]  
Deans D. R., 1968, CHROMATOGRAPHIA, V1, P18, DOI DOI 10.1007/BF02259005
[6]   Oil spill source identification by comprehensive two-dimensional gas chromatography [J].
Gaines, RB ;
Frysinger, GS ;
Hendrick-Smith, MS ;
Stuart, JD .
ENVIRONMENTAL SCIENCE & TECHNOLOGY, 1999, 33 (12) :2106-2112
[7]   INFLUENCE OF INSTRUMENTAL CONTRIBUTIONS ON APPARENT COLUMN EFFICIENCY IN HIGH-SPEED GAS-CHROMATOGRAPHY [J].
GASPAR, G ;
ANNINO, R ;
VIDALMADJAR, C ;
GUIOCHON, G .
ANALYTICAL CHEMISTRY, 1978, 50 (11) :1512-1518
[8]  
Giddings J.C., 1991, Unified separation science
[9]  
Kinghorn RM, 2000, HRC-J HIGH RES CHROM, V23, P245
[10]  
Ledford EB, 2000, HRC-J HIGH RES CHROM, V23, P202