Flow-switching device for comprehensive two-dimensional gas chromatography

被引:84
作者
Bueno, PA [1 ]
Seeley, JV [1 ]
机构
[1] Oakland Univ, Dept Chem, Rochester, MI 48309 USA
关键词
gas chromatography; comprehensive two-dimensional; flow-switching instrumentation; diesel fuel; volatile organic compounds;
D O I
10.1016/j.chroma.2003.10.033
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A simple flow-switching device has been developed as a differential flow modulator for comprehensive two-dimensional gas chromatography (GC x GC). The device is assembled from tubing, four tee unions, and a solenoid valve. The solenoid valve is located outside the oven of the gas chromatograph and is not in the sample path. The modulation technique has no inherent temperature restrictions and passes 100% of the primary column effluent to the secondary column(s). Secondary peaks are produced with widths at half maximum less than 100 ms when operating in GC x 2GC mode with a 2.0 s modulation period. The efficacy of this approach is demonstrated through the analysis of a standard mixture of volatile organic compounds (VOCs) and diesel fuel. (C) 2003 Elsevier B.V. All rights reserved.
引用
收藏
页码:3 / 10
页数:8
相关论文
共 17 条
[11]   COMPREHENSIVE 2-DIMENSIONAL GAS-CHROMATOGRAPHY USING AN ON-COLUMN THERMAL MODULATOR INTERFACE [J].
LIU, ZY ;
PHILLIPS, JB .
JOURNAL OF CHROMATOGRAPHIC SCIENCE, 1991, 29 (06) :227-231
[12]  
SACKS R, 1994, ENVIRON SCI TECHNOL, V28, pA428, DOI 10.1021/es00058a721
[13]  
Seeley JV, 2001, J SEP SCI, V24, P444, DOI 10.1002/1615-9314(20010601)24:6<444::AID-JSSC444>3.0.CO
[14]  
2-5
[15]   Theoretical study of incomplete sampling of the first dimension in comprehensive two-dimensional chromatography [J].
Seeley, JV .
JOURNAL OF CHROMATOGRAPHY A, 2002, 962 (1-2) :21-27
[16]   Comprehensive two-dimensional gas chromatography via differential flow modulation [J].
Seeley, JV ;
Kramp, F ;
Hicks, CJ .
ANALYTICAL CHEMISTRY, 2000, 72 (18) :4346-4352
[17]   Comprehensive two-dimensional gas chromatography of volatile and semi-volatile components using a diaphragm valve-based instrument [J].
Sinha, AE ;
Johnson, KJ ;
Prazen, BJ ;
Lucas, SV ;
Fraga, CG ;
Synovec, RE .
JOURNAL OF CHROMATOGRAPHY A, 2003, 983 (1-2) :195-204