Validation of the conductance catheter method for measurement of ventricular volumes under varying conditions relevant to cardiac surgery

被引:12
作者
Al-Khalidi, AH [1 ]
Townend, JN
Bonser, RS
Coote, JH
机构
[1] Univ Birmingham, Sch Med, Dept Physiol, Birmingham B15 2TT, W Midlands, England
[2] Queen Elizabeth Hosp, Dept Surg, Birmingham B15 2TH, W Midlands, England
[3] Queen Elizabeth Hosp, Dept Cardiovasc Med, Birmingham B15 2TH, W Midlands, England
关键词
D O I
10.1016/S0002-9149(98)00613-4
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The development of the conductance catheter method has enabled continuous measurement of intraventricular volume in vivo, thus making assessment of pump performance of the heart in vivo possible using pressure-volume analysis. However, this method has not been validated under conditions where pump rate, conductance, viscosity, and temperature of the fluid in the heart chamber is changed as happens in cardiac surgery. To validate the method, pressure-volume data were measured by conductance catheter in a physical model of the human left ventricle. The volume, salinity, viscosity, and temperature of the fluid inside the model were rigorously controlled. The measured pressure-volume data were compared with the actual values to assess the accuracy and dependence of the conductance-measured volumes an salinity, viscosity, temperature, and pump rate. Conductance-measured volumes were not significantly different over a range of heart rates extending from 60 to 100 beats/min, and they were not significantly different over a salinity range of 0.2 to 2 normal saline, a viscosity range of 2.7 to 3.5 centipoise, or over a temperature range of 20 degrees C to 39 degrees C. The percentage errors between actual stroke volumes and conductance-derived volumes were of the order of 10% on overage. Our data shows that for a broad range of conditions normally encountered during cardiac surgery, there is no dependence of conductance-measured volume on heart rate, viscosity, temperature, or salinity, provided the correct value of fluid resistivity is used. (C) 1998 by Excerpta Medica, Inc.
引用
收藏
页码:1248 / 1252
页数:5
相关论文
共 15 条
[1]   SIMULTANEOUS CONDUCTANCE CATHETER AND DIMENSION ASSESSMENT OF LEFT-VENTRICLE VOLUME IN THE INTACT ANIMAL [J].
APPLEGATE, RJ ;
CHENG, CP ;
LITTLE, WC .
CIRCULATION, 1990, 81 (02) :638-648
[2]  
Baan J., 1989, Automedica, V11, P357
[3]   CONTINUOUS STROKE VOLUME AND CARDIAC-OUTPUT FROM INTRAVENTRICULAR DIMENSIONS OBTAINED WITH IMPEDANCE CATHETER [J].
BAAN, J ;
JONG, TTA ;
KERKHOF, PLM ;
MOENE, RJ ;
VANDIJK, AD ;
VANDERVELDE, ET ;
KOOPS, J .
CARDIOVASCULAR RESEARCH, 1981, 15 (06) :328-334
[4]   CONTRACTILE STATE OF LEFT-VENTRICLE IN MAN AS EVALUATED FROM END-SYSTOLIC PRESSURE-VOLUME RELATIONS [J].
GROSSMAN, W ;
BRAUNWALD, E ;
MANN, T ;
MCLAURIN, LP ;
GREEN, LH .
CIRCULATION, 1977, 56 (05) :845-852
[5]   FROM EMAX TO PRESSURE-VOLUME RELATIONS - A BROADER VIEW [J].
KASS, DA ;
MAUGHAN, WL .
CIRCULATION, 1988, 77 (06) :1203-1212
[6]   DETERMINATION OF LEFT-VENTRICULAR END-SYSTOLIC PRESSURE-VOLUME RELATIONSHIPS BY THE CONDUCTANCE (VOLUME) CATHETER TECHNIQUE [J].
KASS, DA ;
YAMAZAKI, T ;
BURKHOFF, D ;
MAUGHAN, WL ;
SAGAWA, K .
CIRCULATION, 1986, 73 (03) :586-595
[7]   COMPARATIVE INFLUENCE OF LOAD VERSUS INOTROPIC STATES ON INDEXES OF VENTRICULAR CONTRACTILITY - EXPERIMENTAL AND THEORETICAL-ANALYSIS BASED ON PRESSURE-VOLUME RELATIONSHIPS [J].
KASS, DA ;
MAUGHAN, WL ;
GUO, ZM ;
KONO, A ;
SUNAGAWA, K ;
SAGAWA, K .
CIRCULATION, 1987, 76 (06) :1422-1436
[8]  
PARKER RE, 1979, INTRO STAT BIOL, P75
[9]   THE END-SYSTOLIC PRESSURE-VOLUME RELATION OF THE VENTRICLE - DEFINITION, MODIFICATIONS AND CLINICAL USE [J].
SAGAWA, K .
CIRCULATION, 1981, 63 (06) :1223-1227
[10]   CONTINUOUS MEASUREMENT OF THE PRESSURE VOLUME RELATIONSHIP IN EXPERIMENTAL HEART-FAILURE PRODUCED BY RAPID VENTRICULAR PACING IN CONSCIOUS DOGS [J].
SASAYAMA, S ;
ASANOI, H ;
ISHIZAKA, S .
EUROPEAN HEART JOURNAL, 1992, 13 :47-51