Development of a quantitative relationship between inhibition percentage and both incubation time and inhibitor concentration for inhibition biosensors - theoretical and practical considerations

被引:58
作者
Zhang, S [1 ]
Zhao, H [1 ]
John, R [1 ]
机构
[1] Griffith Univ, Sch Environm & Appl Sci, Gold Coast, Qld 9726, Australia
关键词
biosensors; microelectrode array; inhibition; paraoxon; pesticides; acetylcholinesterase; choline oxidase;
D O I
10.1016/S0956-5663(01)00240-8
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Theoretical and practical insights into the design and development of immobilised enzyme inhibition biosensors are reported. A general mathematical expression relating the percent of enzyme inhibition (i.e. the analytical signal) to both the inhibitor concentration and the incubation time is presented. The relevant physical, chemical and biochemical parameters required by the model are developed and discussed in terms of the inhibition of acetylcholinesterase by the organophosphorous pesticide, paraoxon. A second enzyme, choline oxidase and an amperometric transducer are used to facilitate the determination acetylcholinesterase inhibitor. (C) 2001 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:1119 / 1126
页数:8
相关论文
共 25 条
[1]  
BARCELO D, 1997, ANAL CHIM ACTA, V38, P3
[2]   ANTICHOLINESTERASE ACTIVITY MEASUREMENT BY A CHOLINE BIOSENSOR - APPLICATION IN WATER ANALYSIS [J].
BERNABEI, M ;
CHIAVARINI, S ;
CREMISINI, C ;
PALLESCHI, G .
BIOSENSORS & BIOELECTRONICS, 1993, 8 (05) :265-271
[3]   EVALUATION OF THE USE OF FREE AND IMMOBILIZED ACETYLCHOLINESTERASE FOR PARAOXON DETECTION WITH AN AMPEROMETRIC CHOLINE OXIDASE BASED BIOSENSOR [J].
CREMISINI, C ;
DISARIO, S ;
MELA, J ;
PILLOTON, R ;
PALLESCHI, G .
ANALYTICA CHIMICA ACTA, 1995, 311 (03) :273-280
[4]   DETECTION OF ORGANOPHOSPHOROUS PESTICIDES WITH AN IMMOBILIZED CHOLINESTERASE ELECTRODE [J].
DURAND, P ;
NICAUD, JM ;
MALLEVIALLE, J .
JOURNAL OF ANALYTICAL TOXICOLOGY, 1984, 8 (03) :112-117
[5]   Amperometric flow-through biosensor for the determination of cholinesterase inhibitors [J].
Evtugyn, GA ;
Ivanov, AN ;
Gogol, EV ;
Marty, JL ;
Budnikov, HC .
ANALYTICA CHIMICA ACTA, 1999, 385 (1-3) :13-21
[6]   Sensitivity and selectivity of electrochemical enzyme sensors for inhibitor determination [J].
Evtugyn, GA ;
Budnikov, HC ;
Nikolskaya, EB .
TALANTA, 1998, 46 (04) :465-484
[7]   Increased paraoxon detection with solvents using acetylcholinesterase inactivation measured with a choline oxidase biosensor [J].
Fennouh, S ;
Casimiri, V ;
Burstein, C .
BIOSENSORS & BIOELECTRONICS, 1997, 12 (02) :97-104
[8]   The response of screen-printed enzyme electrodes containing cholinesterases to organo-phosphates in solution and from commercial formulations [J].
Hart, AL ;
Collier, WA ;
Janssen, D .
BIOSENSORS & BIOELECTRONICS, 1997, 12 (07) :645-654
[9]  
HINTSCHE R, 1989, ANAL LETT, V22, P2175
[10]   Detection of paraoxon by continuous flow system based enzyme sensor [J].
Jeanty, G ;
Marty, JL .
BIOSENSORS & BIOELECTRONICS, 1998, 13 (02) :213-218