Performance of a single photon counting microstrip detector for strip pitches down to 10 μm

被引:45
作者
Bergamaschi, A. [1 ]
Broennimann, Ch. [1 ]
Dinapoli, R. [1 ]
Eikenberry, E. [1 ]
Gozzo, F. [1 ]
Henrich, B. [1 ]
Kobas, M. [1 ]
Kraft, P. [1 ]
Patterson, B. [1 ]
Schmitt, B. [1 ]
机构
[1] Paul Scherrer Inst, CH-5232 Villigen, Switzerland
关键词
silicon microstrip detector; photon counting; charge sharing;
D O I
10.1016/j.nima.2008.03.048
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The MYTHEN detector is a one-dimensional microstrip detector with single photon counting readout optimized for time resolved powder diffraction experiments at the Swiss Light Source (SLS). The system has been successfully tested for many different synchrotron radiation applications including phase contrast and tomographic imaging, small angle scattering, diffraction and time resolved pump and probe experiments for X-ray energies down to 5 keV and counting rate up to 3 MHz. The frontend electronics is designed in order to be coupled to 50 mu m pitch microstrip sensors but some interest in enhancing the spatial resolution is arising for imaging and powder diffraction experiments. A test structure with strip pitches in the range 10-50 mu m has been tested and the gain and noise on the readout electronics have been measured for the different strip pitches, observing no large difference down to 25 mu m. Moreover, the effect of the charge sharing between neighboring strips on the spatial resolution has been quantified by measuring the Point Spread Function (PSF) of the system for the different pitches. (c) 2008 Elsevier B.V. All rights reserved.
引用
收藏
页码:163 / 166
页数:4
相关论文
共 3 条
[1]   Clinical mammography at the SYRMEP beam line: Toward the digital detection system [J].
Dreossi, D. ;
Bergamaschi, A. ;
Schmitt, B. ;
Vallazza, E. ;
Arfelli, F. ;
Longo, R. ;
Menk, R. H. ;
Rigon, L. ;
Rokvic, T. ;
Venanzi, C. ;
Castelli, E. .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2007, 576 (01) :160-163
[2]   A method for measuring the presampled MTF of digital radiographic systems using an edge test device [J].
Samei, E ;
Flynn, MJ ;
Reimann, DA .
MEDICAL PHYSICS, 1998, 25 (01) :102-113
[3]   Mythen detector system [J].
Schmitt, B ;
Brönnimann, C ;
Eikenberry, EF ;
Gozzo, F ;
Hörmann, C ;
Horisberger, R ;
Patterson, B .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2003, 501 (01) :267-272