Performance evaluation of a "dual-side read" dedicated mammography computed radiography system

被引:38
作者
Fetterly, KA [1 ]
Schueler, BA [1 ]
机构
[1] Mayo Clin, Rochester, MN 55905 USA
关键词
mammography; digital; computed radiography; MTF; NPS; DQE;
D O I
10.1118/1.1584045
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The image quality of a dedicated mammography computed radiography (CR) system was characterized. A unique feature of this system is that it collects image signals from both sides of the storage phosphor. Measurements of the modulation transfer function (MTF), noise power spectrum (NPS), and detective quantum efficiency (DQE) were made. This work included improvements in our measurement methods to specifically account for the detrimental effects of system glare on the MTF and to accurately characterize the low-frequency NPS components. linage quality measurements were performed using a 25 kVp beam filtered with 2 mm Al and an exposure range of 1 to 100 mR (87 to 870 muGy). The DQE was found to decrease with increasing exposure due to an increased contribution of storage phosphor structure noise. The DQE of this system was compared to similar measurements made using a standard CR system. The dual-side read system demonstrated superior DQE compared to the standard system. The decrease in DQE with increasing exposure was more severe for the standard system than the dual-side read system. This finding suggests that the CR system noise was reduced for the dual-side read system compared to the standard system. (C) 2003 American Association of Physicists in Medicine.
引用
收藏
页码:1843 / 1854
页数:12
相关论文
共 34 条
[1]  
*AM COLL RAD, 1999, AM COLL RAD MAMM QUA
[2]   Improvement of image quality in CR mammography by detection of emissions from dual sides of an imaging plate [J].
Arakawa, S ;
Yasuda, H ;
Kohda, K ;
Suzuki, T .
MEDICAL IMAGING 2000: PHYSICS OF MEDICAL IMAGING, 2000, 3977 :590-600
[3]   Molybdenum, rhodium, and tungsten anode spectral models using interpolating polynomials with application to mammography [J].
Boone, JM ;
Fewell, TR ;
Jennings, RJ .
MEDICAL PHYSICS, 1997, 24 (12) :1863-1874
[4]   Spectral modeling and compilation of quantum fluence in radiography and mammography [J].
Boone, JM .
PHYSICS OF MEDICAL IMAGING, 1998, 3336 :592-601
[5]   Comparison of x-ray cross sections for diagnostic and therapeutic medical physics [J].
Boone, JM ;
Chavez, AE .
MEDICAL PHYSICS, 1996, 23 (12) :1997-2005
[6]   Advances in high-speed mammographic image quality [J].
Bunch, PC .
MEDICAL IMAGING 1999: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2, 1999, 3659 :120-130
[7]   PHYSICAL ASPECTS OF PHOTOSTIMULABLE PHOSPHOR COMPUTED RADIOGRAPHY [J].
COWEN, AR ;
WORKMAN, A ;
PRICE, JS .
BRITISH JOURNAL OF RADIOLOGY, 1993, 66 (784) :332-345
[8]   A METHOD FOR MODULATION TRANSFER-FUNCTION DETERMINATION FROM EDGE PROFILES WITH CORRECTION FOR FINITE-ELEMENT DIFFERENTIATION [J].
CUNNINGHAM, IA ;
FENSTER, A .
MEDICAL PHYSICS, 1987, 14 (04) :533-537
[9]  
CUNNINGHAM IA, 1999, PRACTICAL DIGITAL IM, pCH8
[10]   Image-quality performance of an a-Si: H-based X-ray imaging system for digital mammography [J].
Darambara, DG ;
Taibi, A ;
Speller, RD .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2002, 477 (1-3) :521-526