Direct comparison of 3D spiral vs. Cartesian gradient-echo coronary magnetic resonance angiography

被引:47
作者
Börnert, P
Stuber, M
Botnar, RM
Kissinger, KV
Koken, P
Spuentrup, E
Manning, WJ
机构
[1] Philips Res Labs, D-22335 Hamburg, Germany
[2] Beth Israel Deaconess Med Ctr, Boston, MA 02215 USA
[3] Harvard Univ, Sch Med, Boston, MA USA
[4] Philips Med Syst, Best, Netherlands
关键词
coronary angiography; MRI; 3D imaging; spiral imaging; Cartesian imaging;
D O I
10.1002/mrm.1258
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
While 3D thin-slab coronary magnetic resonance angiography (MRA) has traditionally been performed using a Cartesian acquisition scheme, spiral k-space data acquisition offers several potential advantages. However, these strategies have not been directly compared in the same subjects using similar methodologies. Thus, in the present study a comparison was made between 3D coronary MRA using Cartesian segmented k-space gradient-echo and spiral k-space data acquisition schemes. In both approaches the same spatial resolution was used and data were acquired during free breathing using navigator gating and prospective slice tracking. Magnetization preparation (T-2 preparation and fat suppression) was applied to increase the contrast. For spiral imaging two different examinations were performed, using one or two spiral interleaves, during each R-R Interval. Spiral acquisitions were found to be superior to the Cartesian scheme with respect to the signal-to-noise ratio (SNR) and contrast-to-noise-ratio (CNR) (both P < 0.001) and image quality. The single spiral per R-R interval acquisition had the same total scan duration as the Cartesian acquisition, but the single spiral had the best image quality and a 2.6-fold increase in SNR. The double-interleaf spiral approach showed a 50% reduction in scanning time, a 1.8-fold increase in SNR, and similar image quality when compared to the standard Cartesian approach. Spiral 3D coronary MRA appears to be preferable to the Cartesian scheme. The increase in SNR may be "traded" for either shorter scanning times using multiple consecutive spiral interleaves, or for enhanced spatial resolution. (C) 2001 Wiley-Liss, Inc.
引用
收藏
页码:789 / 794
页数:6
相关论文
共 30 条
  • [1] CORONARY-ARTERY IMAGING AT 0.5 T USING SEGMENTED 3D ECHO-PLANAR IMAGING
    BORNERT, P
    JENSEN, D
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1995, 34 (06) : 779 - 785
  • [2] Börnert P, 2001, MAGNET RESON MED, V45, P172, DOI 10.1002/1522-2594(200101)45:1<172::AID-MRM1024>3.0.CO
  • [3] 2-L
  • [4] Botnar RM, 1999, JMRI-J MAGN RESON IM, V10, P821, DOI 10.1002/(SICI)1522-2586(199911)10:5<821::AID-JMRI29>3.0.CO
  • [5] 2-7
  • [6] Improved coronary artery definition with T2-weighted, free-breathing, three-dimensional coronary MRA
    Botnar, RM
    Stuber, M
    Danias, PG
    Kissinger, KV
    Manning, WJ
    [J]. CIRCULATION, 1999, 99 (24) : 3139 - 3148
  • [7] CORONARY ANGIOGRAPHY WITH MAGNETIZATION-PREPARED T-2 CONTRAST
    BRITTAIN, JH
    HU, BS
    WRIGHT, GA
    MEYER, CH
    MACOVSKI, A
    NISHIMURA, DG
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1995, 33 (05) : 689 - 696
  • [8] ADAPTIVE TECHNIQUE FOR HIGH-DEFINITION MR IMAGING OF MOVING STRUCTURES
    EHMAN, RL
    FELMLEE, JP
    [J]. RADIOLOGY, 1989, 173 (01) : 255 - 263
  • [9] Gatehouse PD, 1999, MAGNET RESON MED, V41, P1023, DOI 10.1002/(SICI)1522-2594(199905)41:5<1023::AID-MRM22>3.0.CO
  • [10] 2-1