The peripheral pulse wave: Information overlooked

被引:141
作者
Murray, WB [1 ]
Foster, PA [1 ]
机构
[1] PENN STATE UNIV,MILTON S HERSHEY MED CTR,COLL MED,DEPT ANESTHESIA,HERSHEY,PA 17033
来源
JOURNAL OF CLINICAL MONITORING | 1996年 / 12卷 / 05期
关键词
pulse oximetry; peripheral pulse wave; arterial pulse wave; hypovolemia; Valsalva maneuver; dicrotic notch; plethysmography;
D O I
10.1007/BF02077634
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Although the waveform derived from a peripheral pulse monitor or pulse oximeter may resemble an arterial pressure waveform, it is in fact a visualization of blood volume change in transilluminated tissue caused by passage of blood: an indication of perfusion or blood flow. Most currently available pulse oximeters indicate this flow, but few display it in usable form. Since adequate tissue blood flow is a prerequisite for normal metabolic activity it is a parameter that should merit a place in standard anesthesia or intensive care monitors. That the peripheral tissue blood flow is not routinely displayed may be in part due to the difficulty in quantifying data obtained: flow is not accurately measured as simply as pressure, even by invasive means. It is in the pattern of the waveform that beat-to-beat changes in stroke volume can be better seen than measured, or in the interaction of ventilation and circulation that tests general circulatory performance. The origin and interpretation of these changes are discussed and illustrated with examples. We indicate how new physiological tests of autonomic function and cardiac preload can be developed using pulse plethysmography. The importance and application of the Valsalva effect on the waveform is emphasized. This effect is particularly applicable for monitoring adequate fluid loading and the action of vasodilator drugs, which are both important in anesthesia. Differences between the arterial pulse pressure wave and tissue flow wave are discussed, as well as the cause of certain artifacts, including the wandering dicrotic notch.
引用
收藏
页码:365 / 377
页数:13
相关论文
共 33 条
[1]  
BROOME IJ, 1988, ANAESTHESIA, V43, P833
[2]  
BRUNER J, 1978, HDB BLOOD PRESSURE M
[3]   PULSE OXIMETER PROBES - A COMPARISON BETWEEN FINGER, NOSE, EAR AND FOREHEAD PROBES UNDER CONDITIONS OF POOR PERFUSION [J].
CLAYTON, DG ;
WEBB, RK ;
RALSTON, AC ;
DUTHIE, D ;
RUNCIMAN, WB .
ANAESTHESIA, 1991, 46 (04) :260-265
[4]  
DALY M, 1986, HDB PHYSL RESP SYSTE, P569
[5]  
DORLAS JC, 1986, CLOSED CIRCUIT SYSTE, P136
[6]   HEART RATE RESPONSE TO VALSALVA MANEUVER AS A TEST OF CIRCULATORY INTEGRITY [J].
ELISBERG, EI .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1963, 186 (03) :200-&
[7]  
FREEMAN DH, 1994, DISCOVERY, V15, P62
[8]   DIRECT BLOOD-PRESSURE MEASUREMENT - DYNAMIC-RESPONSE REQUIREMENTS [J].
GARDNER, RM .
ANESTHESIOLOGY, 1981, 54 (03) :227-236
[9]  
GUYTON AC, 1991, TXB PHYSL, P160
[10]   Physiologic relationships between intrathoracic, intraspinal and arterial pressures [J].
Hamilton, WF ;
Woodbury, RA ;
Harper, HT .
JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1936, 107 :853-856