Direct traceability of the portable QASUME irradiance scale to the primary irradiance standard of the PTB

被引:35
作者
Gröbner, J
Sperfeld, P
机构
[1] Commiss European Communities, Joint Res Ctr, Inst Hlth & Consumer Protect, I-21020 Ispra, Italy
[2] Phys Tech Bundesanstalt, D-3300 Braunschweig, Germany
关键词
D O I
10.1088/0026-1394/42/2/008
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The portable Quality Assurance of Spectral UV Measurements in Europe (QASUME) irradiance scale has been directly compared to the primary irradiance standard (blackbody BB3200pg) of the Physikalisch-Technische Bundesanstalt (PTB) using the travelling spectroradiometer QASUME. The average blackbody spectral irradiance scale based on three different temperatures (2940 K, 2997 K and 3059 K) was on average 0.7% greater than the QASUME irradiance scale based on the average of three low-power lamps. This is fully consistent with the expanded uncertainty (k = 2) of 4% in the ultraviolet (UV) wavelength range and 2.6% above 400 nm of the QASUME irradiance scale. If these blackbody measurements are used to establish directly the portable QASUME irradiance scale, then its expanded uncertainty (k = 2) is substantially reduced to 2.5% in the UV and 2.0% above 400 nm.
引用
收藏
页码:134 / 139
页数:6
相关论文
共 7 条
[1]  
GROBNER J, 2003, B5503 EUR EN
[2]  
GROBNER J, 2004, P 20 QUAD OZ S KOS G
[3]  
SECKMEYER G, 2001, 12K WMO GAW
[4]   Improvement and extension of the black-body-based spectral irradiance scale [J].
Sperfeld, P ;
Metzdorf, J ;
Yousef, SG ;
Stock, KD ;
Möller, W .
METROLOGIA, 1998, 35 (04) :267-271
[5]  
SPERFELD P, 1999, ENTWICKLUNG EINER EM
[6]  
World Health Organization [WHO], 1994, ENV HLTH CRIT EHC
[7]  
2005, IN PRESS FINAL REP S