Dynamic range requirements for MRI

被引:31
作者
Behin, R
Bishop, J
Henkelman, RM
机构
[1] Hosp Sick Children, Mouse Imaging Ctr, Toronto, ON M5G 1X8, Canada
[2] Univ Toronto, Dept Med Biophys, Toronto, ON, Canada
关键词
MRI; high resolution; dynamic range; analog to digital; quantization noise; thermal noise; variable gain; signal-to-noise ratio (SNR); k-space sampling;
D O I
10.1002/cmr.b.20042
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
in order to realize the full potential in an MR receiver, the digitizer must capture a signal magnitude range from the central k-space peak to the thermal noise floor of the system. This dynamic range can exceed the performance of standard 16-bit data converters. For example, a whole-body mouse scan in a 7 Tesla magnet requires 20 bits of dynamic range. A 3D high-resolution mouse scan at 75 mu m isotropic voxel resolution using a 16-bit spectrometer shows an eightfold improvement in image SNR by gain stepping the receiver prior to digitization to cover the full magnitude dynamic range compared to a standard fixed gain approach. A method is presented to determine the dynamic range requirements of any experiment. (c) 2005 Wiley Periodicals, Inc.
引用
收藏
页码:28 / 35
页数:8
相关论文
共 10 条
[1]  
[Anonymous], P INT SOC MAGNETIC R
[2]   IMPROVEMENT OF DYNAMIC-RANGE IN NMR BY OVERSAMPLING [J].
DELSUC, MA ;
LALLEMAND, JY .
JOURNAL OF MAGNETIC RESONANCE, 1986, 69 (03) :504-507
[3]   Improved resolution and signal-to-noise ratio in MRI via enhanced signal digitization [J].
Elliott, MA ;
Insko, EK ;
Greenman, RL ;
Leigh, JS .
JOURNAL OF MAGNETIC RESONANCE, 1998, 130 (02) :300-304
[4]   RELATIONS BETWEEN THE STATISTICS OF NATURAL IMAGES AND THE RESPONSE PROPERTIES OF CORTICAL-CELLS [J].
FIELD, DJ .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1987, 4 (12) :2379-2394
[5]   THE INFORMATION-CONTENT OF MR IMAGES [J].
FUDERER, M .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1988, 7 (04) :368-380
[6]   MEASUREMENT OF SIGNAL INTENSITIES IN THE PRESENCE OF NOISE IN MR IMAGES [J].
HENKELMAN, RM .
MEDICAL PHYSICS, 1985, 12 (02) :232-233
[7]   k-space in the clinic [J].
Paschal, CB ;
Morris, HD .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2004, 19 (02) :145-159
[8]   STATISTICS OF NATURAL IMAGES - SCALING IN THE WOODS [J].
RUDERMAN, DL ;
BIALEK, W .
PHYSICAL REVIEW LETTERS, 1994, 73 (06) :814-817
[9]   k-space interpretation of the Rose Model:: Noise limitation on the detectable resolution in MRI [J].
Watts, R ;
Wang, Y .
MAGNETIC RESONANCE IN MEDICINE, 2002, 48 (03) :550-554
[10]   DYNAMIC-RANGE COMPRESSION IN MRI BY MEANS OF A NONLINEAR GRADIENT PULSE [J].
WEDEEN, VJ ;
CHAO, YS ;
ACKERMAN, JL .
MAGNETIC RESONANCE IN MEDICINE, 1988, 6 (03) :287-295