Evaluation of a new generation of plastic evacuated blood-collection tubes in clinical chemistry, therapeutic drug monitoring, hormone and trace metal analysis

被引:30
作者
Boeynaems, JM [1 ]
De Leener, A [1 ]
Dessars, B [1 ]
Villa-Lobos, HR [1 ]
Aubry, JC [1 ]
Cotton, F [1 ]
Thiry, P [1 ]
机构
[1] Free Univ Brussels, Hop Erasme, Med Chem Lab, Dept Clin Pathol, B-1070 Brussels, Belgium
关键词
evacuated tube; polyethylene terephthalate;
D O I
10.1515/CCLM.2004.013
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Polyethylene terephthalate (PET) tubes have several advantages over glass tubes: they are unbreakable, lighter and more easily disposed of. Despite a steady increase in their use and an expansion of the range of available tubes, few studies validating their use have been published in the literature. This paper describes the various studies that have been performed to compare VENOJECT glass, VENOSAFE PET and VENOSAFE PET/heparin tubes for the assay of a panel of analytes in routine clinical chemistry, immunochemistry, hormone and tumor marker analysis and trace metal determination. These studies demonstrate that VENOSAFE PET tubes are a suitable alternative to glass tubes.
引用
收藏
页码:67 / 71
页数:5
相关论文
共 15 条
[1]   Evaluation of polyethylene terephthalate for ABO and Rh typing and alloantibody screening [J].
Anderson, DR ;
Wiseman, J ;
MacLeod, J ;
Burton, E ;
Zayed, E .
TRANSFUSION, 2000, 40 (06) :669-672
[2]   Heavy metal poisoning and its laboratory investigation [J].
Baldwin, DR ;
Marshall, WJ .
ANNALS OF CLINICAL BIOCHEMISTRY, 1999, 36 :267-300
[3]  
*BIOTOX, 2000, INV LAB EFF DOS BIOL
[4]  
Burtis C., 1999, TIETZ TXB CLIN CHEM
[5]   Stability of therapeutic drug measurement in specimens collected in VACUTAINER plastic blood-collection tubes [J].
Dasgupta, A ;
Blackwell, W ;
Bard, D .
THERAPEUTIC DRUG MONITORING, 1996, 18 (03) :306-309
[6]  
Esernio-Jenssen D, 1999, CLIN CHEM, V45, P148
[7]  
HILL BM, 1992, CLIN CHEM, V38, P1474
[8]  
JUPPNER H, 1980, J CLIN CHEM CLIN BIO, V18, P585
[9]   A NEW PLASTIC EVACUATED TUBE WITH PLASMA SEPARATOR [J].
LANDT, M ;
WILHITE, TR ;
SMITH, CH .
JOURNAL OF CLINICAL LABORATORY ANALYSIS, 1995, 9 (02) :101-106
[10]  
LINDSTEDT G, 1985, CLIN CHEM, V31, P76