A phased array echoplanar imaging system for fMRI

被引:8
作者
Frederick, BD [1 ]
Wald, LL [1 ]
Maas, LC [1 ]
Renshaw, PF [1 ]
机构
[1] McLean Hosp, Brain Imaging Ctr, Belmont, MA 02478 USA
关键词
phased array; echoplanar imaging; functional imaging; BOLD;
D O I
10.1016/S0730-725X(98)00157-X
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A fully parallel, simultaneous sampling phased array receiver system for a clinical echoplanar imaging system is described and evaluated for BOLD activation and relative cerebral blood volume (rCBV) experiments. A 4-coil array curved around the occipital lobe improved SNR by factors of 1.5 in the visual cortex and 3.1 in the visual association cortex relative to a 13-cm diameter surface coil, improving the statistical significance and coverage of visual activation maps. A I-coil bilateral array increased SNR throughout the head relative to a quadrature head coil by up to a factor of 5 in much of the cortex, with proportional improvement in the SNR of rCBV maps. (C) 1998 Elsevier Science Inc.
引用
收藏
页码:121 / 129
页数:9
相关论文
共 20 条
  • [1] BANDETTINI P, 1993, MAGNET RESON MED, V20, P161
  • [2] FUNCTIONAL CEREBRAL IMAGING BY SUSCEPTIBILITY-CONTRAST NMR
    BELLIVEAU, JW
    ROSEN, BR
    KANTOR, HL
    RZEDZIAN, RR
    KENNEDY, DN
    MCKINSTRY, RC
    VEVEA, JM
    COHEN, MS
    PYKETT, IL
    BRADY, TJ
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1990, 14 (03) : 538 - 546
  • [3] Signal-to-noise analysis of cerebral blood volume maps from dynamic NMR imaging studies
    Boxerman, JL
    Rosen, BR
    Weisskoff, RM
    [J]. JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1997, 7 (03): : 528 - 537
  • [4] Phase alignment of multiple surface coil data for reduced bandwidth and reconstruction requirements
    Debbins, JP
    Felmlee, JP
    Riederer, SJ
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1997, 38 (06) : 1003 - 1011
  • [5] Harris GJ, 1996, AM J PSYCHIAT, V153, P721
  • [6] TEMPORAL LOBES - SURFACE MR COIL PHASED-ARRAY IMAGING
    HAYES, CE
    TSURUDA, JS
    MATHIS, CM
    [J]. RADIOLOGY, 1993, 189 (03) : 918 - 920
  • [7] VOLUME IMAGING WITH MR PHASED-ARRAYS
    HAYES, CE
    HATTES, N
    ROEMER, PB
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1991, 18 (02) : 309 - 319
  • [8] Functional magnetic resonance imaging of the human brain
    Kim, SG
    Ugurbil, K
    [J]. JOURNAL OF NEUROSCIENCE METHODS, 1997, 74 (02) : 229 - 243
  • [9] DYNAMIC MAGNETIC-RESONANCE-IMAGING OF HUMAN BRAIN ACTIVITY DURING PRIMARY SENSORY STIMULATION
    KWONG, KK
    BELLIVEAU, JW
    CHESLER, DA
    GOLDBERG, IE
    WEISSKOFF, RM
    PONCELET, BP
    KENNEDY, DN
    HOPPEL, BE
    COHEN, MS
    TURNER, R
    CHENG, HM
    BRADY, TJ
    ROSEN, BR
    [J]. PROCEEDINGS OF THE NATIONAL ACADEMY OF SCIENCES OF THE UNITED STATES OF AMERICA, 1992, 89 (12) : 5675 - 5679
  • [10] Decoupled automated rotational and translational registration for functional MRI time series data: The DART registration algorithm
    Maas, LC
    Frederick, BD
    Renshaw, PF
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1997, 37 (01) : 131 - 139