Improvement in B1-inhomogeneity artifacts in the abdomen at 3T MR imaging using a radiofrequency cushion

被引:56
作者
Franklin, Kendra M. [1 ]
Dale, Brian M. [2 ]
Merkle, Elmar M. [1 ]
机构
[1] Duke Univ, Med Ctr, Dept Radiol, Durham, NC 27710 USA
[2] Siemens Med Solut, Cary, NC USA
关键词
radiofrequency cushion; B1-inhomogeneity artifacts; 3; Tesla; abdominal imaging; dielectric artifacts;
D O I
10.1002/jmri.21164
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 [临床医学]; 100207 [影像医学与核医学]; 1009 [特种医学];
摘要
Purpose: To determine the impact of a radiofrequency (RF) cushion on abdominal image quality at 3T. Materials and Methods: Twenty volunteers were scanned on a 3T system with and without the use of an RF cushion, T2-weighted HASTE and true-FISP sequences were used for coronal imaging, and each sequence was performed with and without integrated parallel acquisition techniques (iPAT). Additionally, axial T2-weighted HASTE and true-FISP imaging without iPAT was performed. The effect on image quality was assessed by subjectively evaluating the presence and severity of RF inhomogeneity artifacts by two independent readers in a side-by-side comparison with a 5-point differential receiver operating characteristics curve. Results: Both readers considered the quality of HASTE images acquired with the RF cushion to be significantly better in all cases. The quality of true-FISP images did not benefit significantly from the use of the RF cushion in any case. Conclusion: The RF cushion significantly improves image quality for abdominal imaging at 3T by reducing or eliminating B1-inhomogenity artifacts. This improved image quality is seen on turbo spin echo-based T2-weighted imaging such as HASTE but does not apply to gradient echo-based T2-weighted imaging such as the true-FISP sequence.
引用
收藏
页码:1443 / 1447
页数:5
相关论文
共 16 条
[1]
Spiral volume coil for improved RF field homogeneity at high static magnetic field strength [J].
Alsop, DC ;
Connick, TJ ;
Mizsei, G .
MAGNETIC RESONANCE IN MEDICINE, 1998, 40 (01) :49-54
[2]
Imaging artifacts at 3.OT [J].
Bernstein, Matt A. ;
Huston, John, III ;
Ward, Heidi A. .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2006, 24 (04) :735-746
[3]
The differential receiver operating characteristic (DROC) method [J].
Chakraborty, DP ;
Kundel, HL ;
Nodine, C ;
Narayan, TK ;
Devaraju, V .
MEDICAL IMAGING 1998: IMAGE PROCESSING, PTS 1 AND 2, 1998, 3338 :234-240
[4]
Central brightening due to constructive interference with, without, and despite dielectric resonance [J].
Collins, CM ;
Liu, WZ ;
Schreiber, W ;
Yang, QX ;
Smith, MB .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2005, 21 (02) :192-196
[5]
The dielectric properties of biological tissues .1. Literature survey [J].
Gabriel, C ;
Gabriel, S ;
Corthout, E .
PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (11) :2231-2249
[6]
HAACKE EM, 1967, MAGNETIC RESONANCE I, P1
[7]
Merkle Elmar M, 2006, Magn Reson Imaging Clin N Am, V14, P17, DOI 10.1016/j.mric.2005.12.001
[8]
Abdominal MRI at 3.0 T: The basics revisited [J].
Merkle, EM ;
Dale, BM .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2006, 186 (06) :1524-1532
[9]
Small tip angle three-dimensional tailored radiofrequency slab-select pulse for reduced B1 inhomogeneity at 3 T [J].
Saekho, S ;
Boada, FE ;
Noll, DC ;
Stenger, VA .
MAGNETIC RESONANCE IN MEDICINE, 2005, 53 (02) :479-484
[10]
Whole-body MRI at high field: technical limits and clinical potential [J].
Schick, F .
EUROPEAN RADIOLOGY, 2005, 15 (05) :946-959