Evaluation procedures for reversed-phase high-performance liquid chromatographic columns in the analysis of strongly basic compounds using principal components analysis for data assessment

被引:32
作者
Brereton, RG
McCalley, DV
机构
[1] Univ Bristol, Sch Chem, Bristol BS8 1TS, Avon, England
[2] Univ W England, Fac Sci Appl, Bristol BS16 1QY, Avon, England
关键词
high-performance liquid chromatography; chemometrics; column assessment; pattern recognition; pharmaceuticals;
D O I
10.1039/a800149i
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The performance of eight high-purity silica-based reversed-phase columns was analysed using chemometric pattern recognition. Tests were performed at pH 3.0 and 7.0, using 10 compounds, three mobile phase modifiers (methanol, tetrahydofuran and acetonitrile) and four parameters (retention factor (k'), column efficiency (N), the Dorsey-Foley efficiency [N(df)] and the asymmetry factor (A(s)])}. Principal components analysis was performed on subsets of the data. Features of both scores and loadings plots, together with the corresponding correlation coefficients between chromatographic tests, are discussed in detail. It is concluded that methanol and acetonitrile exhibit similar properties, the relative performance at each pH differs, the number of compounds used in the test can be reduced to a recommended five, and several of the original tests are redundant, Procedures are recommended for determining a range of tests that provide information on the column performance, which can be applied to other situations.
引用
收藏
页码:1175 / 1185
页数:11
相关论文
共 17 条
[1]  
[Anonymous], 1979, Multivariate analysis
[2]  
Brereton R.G., 1993, CHEMOMETRICS APPL MA
[3]  
BRERETON RG, 1992, MULTIVARIATE PATTERN
[4]   Procrustes analysis for the determination of number of significant masses in gas chromatography mass spectrometry [J].
Demir, C ;
Hindmarch, P ;
Brereton, RG .
ANALYST, 1996, 121 (10) :1443-1449
[5]   COMPARISON AND CHARACTERIZATION OF REVERSED PHASES [J].
ENGELHARDT, H ;
JUNGHEIM, M .
CHROMATOGRAPHIA, 1990, 29 (1-2) :59-68
[6]  
Engelhardt H., 1997, LC GC EUR, V10, P803
[7]   SPECTRAL MAPPING ANALYSIS - A METHOD FOR THE CHARACTERIZATION OF STATIONARY PHASES [J].
HAMOIR, T ;
SANCHEZ, FC ;
BOURGUIGNON, B ;
MASSART, DL .
JOURNAL OF CHROMATOGRAPHIC SCIENCE, 1994, 32 (11) :488-498
[8]   Effect of organic solvent modifier and nature of solute on the performance of bonded silica reversed-phase columns for the analysis of strongly basic compounds by high-performance liquid chromatography [J].
McCalley, DV .
JOURNAL OF CHROMATOGRAPHY A, 1996, 738 (02) :169-179
[9]   INFLUENCE OF ORGANIC-SOLVENT MODIFIER AND SOLVENT STRENGTH ON PEAK SHAPE OF SOME BASIC COMPOUNDS IN HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY USING A REVERSED-PHASE COLUMN [J].
MCCALLEY, DV .
JOURNAL OF CHROMATOGRAPHY A, 1995, 708 (02) :185-194
[10]   Comparative evaluation of bonded-silica reversed-phase columns for high-performance liquid chromatography using strongly basic compounds and alternative organic modifiers buffered at acid pH [J].
McCalley, DV .
JOURNAL OF CHROMATOGRAPHY A, 1997, 769 (02) :169-178