Pace-termination and pacing for prevention of atrial tachyarrhythmias:: Results from a multicenter study with an implantable device for atrial therapy

被引:114
作者
Israel, CW
Hügl, B
Unterberg, C
Lawo, T
Kennis, I
Hettrick, D
Hohnloser, SH
机构
[1] Goethe Univ Frankfurt, Dept Med, Div Cardiol, D-60590 Frankfurt, Germany
[2] Herz Zentrum, Leipzig, Germany
[3] Univ Gottingen Kliniken, Gottingen, Germany
[4] Ruhr Univ Bochum, Dept Cardiol, Bergmannsheil, D-4630 Bochum, Germany
[5] Medtronic Inc, Minneapolis, MN USA
关键词
atrial fibrillation; atrial flutter; antitachycardia pacing; overdrive pacing; dual-chamber pacing; pacemaker algorithm; pacemaker diagnostics;
D O I
10.1046/j.1540-8167.2001.01121.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Pacing for Treatment of Atrial Tachyarrhythmias. Introduction: Patients with bradycardia requiring permanent pacing frequently suffer from additional atrial tachyarrhythmias (ATs). This study evaluated the safety and efficacy of atrial antitachycardia pacing (ATP) and the performance of pacing for AT prevention implemented into a new pacemaker. Methods and Results: In patients with conventional indications for permanent pacing, an investigational DDDRP pacemaker (Medtronic AT500, model 7253) was implanted. The primary study objectives were to determine the safety of overall device functioning and its efficacy in terminating spontaneous AT. A secondary endpoint was to determine the reliability of AT detection. Pacemaker memory functions were used to analyze the impact of dedicated pacing algorithms on AT prevention. In 33 European and Canadian centers, 325 patients were enrolled (mean follow-up 2.3 +/-1.3 months). Complication-free survival at 3 months was 88%. In 2,145 episodes stored with atrial electrograms, AT detection was confirmed in 97%. The algorithm for continuous overdrive pacing increased the percentage of atrial pacing to 97%. After ATP activation, 16,683 of 52,468 AT episodes were treated (120 patients). Of these, 8,903 episodes (53%) were terminated successfully by ATP. No proarrhythmic effect of preventive pacing or atrial ATP was observed. Preventive pacing algorithms increased the median percentage of atrial pacing from 62% to 97%. However, the number of AT/AF (atrial fibrillation) episodes (4.1 vs 4.1 per patient per day) and the time in AT/AF (13.7% vs 12.8%) was not significantly different before and after activation of preventive pacing. Conclusion: DDDRP pacing with a new system for AT therapy was safe and associated with successful pace-termination of AT in 53% of episodes. Preventive pacing and atrial ATP algorithms represent two new functions that can be implemented safely into pacemaker systems for nonpharmacologic treatment of ATs in patients requiring pacemaker therapy.
引用
收藏
页码:1121 / 1128
页数:8
相关论文
共 39 条
  • [1] REGIONAL CONTROL OF ATRIAL-FIBRILLATION BY RAPID PACING IN CONSCIOUS DOGS
    ALLESSIE, M
    KIRCHHOF, C
    SCHEFFER, GJ
    CHORRO, F
    BRUGADA, J
    [J]. CIRCULATION, 1991, 84 (04) : 1689 - 1697
  • [2] INDEPENDENT RISK-FACTORS FOR ATRIAL-FIBRILLATION IN A POPULATION-BASED COHORT - THE FRAMINGHAM HEART-STUDY
    BENJAMIN, EJ
    LEVY, D
    VAZIRI, SM
    DAGOSTINO, RB
    BELANGER, AJ
    WOLF, PA
    [J]. JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1994, 271 (11): : 840 - 844
  • [3] Effective prevention of atrial fibrillation by continuous atrial overdrive pacing after coronary artery bypass surgery
    Blommaert, D
    Gonzalez, M
    Mucumbitsi, J
    Gurné, O
    Evrard, P
    Buche, M
    Louagie, Y
    Eucher, P
    Jamart, J
    Installé, E
    De Roy, L
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2000, 35 (06) : 1411 - 1415
  • [4] PROSPECTIVE EVALUATION OF TRANS-ESOPHAGEAL PACING FOR THE INTERRUPTION OF ATRIAL-FLUTTER
    CRAWFORD, W
    PLUMB, VJ
    EPSTEIN, AE
    KAY, GN
    [J]. AMERICAN JOURNAL OF MEDICINE, 1989, 86 (06) : 663 - 667
  • [5] Prevalence of supraventricular arrhythmias from the automated analysis of data stored in the DDD pacemakers of 617 patients: The AIDA study
    Defaye, P
    Dournaux, F
    Mouton, E
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1998, 21 (01): : 250 - 255
  • [6] Efficacy of intravenous propafenone in termination of atrial flutter by overdrive transesophageal pacing previously ineffective
    DEste, D
    Bertaglia, E
    Mantovan, R
    Zanocco, A
    Franceschi, M
    Pascotto, P
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 1997, 79 (04) : 500 - &
  • [7] Diagnosis and therapy of atrial tachyarrhythmias in the dual chamber implantable cardioverter defibrillator
    Dijkman, B
    Wellens, HJJ
    [J]. JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2000, 11 (11) : 1196 - 1205
  • [8] Effects of biatrial pacing in prevention of postoperative atrial fibrillation after coronary artery bypass surgery
    Fan, K
    Lee, KL
    Chiu, CSW
    Lee, JWT
    He, GW
    Cheung, D
    Sun, MP
    Lau, CP
    [J]. CIRCULATION, 2000, 102 (07) : 755 - 760
  • [9] PREVALENCE, AGE DISTRIBUTION, AND GENDER OF PATIENTS WITH ATRIAL-FIBRILLATION - ANALYSIS AND IMPLICATIONS
    FEINBERG, WM
    BLACKSHEAR, JL
    LAUPACIS, A
    KRONMAL, R
    HART, RG
    [J]. ARCHIVES OF INTERNAL MEDICINE, 1995, 155 (05) : 469 - 473
  • [10] Garrigue S, 1996, ARCH MAL COEUR VAISS, V89, P873