Mechanical performance of ceramic acetabular liners under impact conditions

被引:26
作者
Maher, SA
Lipman, JD
Curley, LJ
Gilchrist, M
Wright, TM
机构
[1] Hosp Special Surg, Lab Biomed Mech & Mat, New York, NY 10021 USA
[2] Univ Coll Dublin, Dept Engn Mech, Dublin 2, Ireland
关键词
ceramic acetabular liner; impingement; impact; total hip arthroplasty; alternate bearings; alumina;
D O I
10.1016/S0883-5403(03)00335-8
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Although new generation alumina ceramics have exhibited a reduced incidence of fracture, concern still persists about the behavior of ceramic acetabular liners under impact conditions. The objective of this study was to explore whether fracture of a new generation alumina ceramic liner was likely to occur in vivo. Ceramic liners were impacted with forces of 23, 21, 15, and 12 kN (n = 3 at each force). At 23 kN, all 3 ceramic liners fractured on the first impact; at 12 kN none of the ceramic liners fractured after 20 impacts. The threshold force of 12 kN is large in comparison with estimated physiologic forces on the hip during falls or stumbling, suggesting that ceramic liner fracture is not a definite consequence of liner impact.
引用
收藏
页码:936 / 941
页数:6
相关论文
共 30 条
[1]
[Anonymous], CLIN ORTHOP
[2]
HIP-JOINT LOADING DURING WALKING AND RUNNING, MEASURED IN 2 PATIENTS [J].
BERGMANN, G ;
GRAICHEN, F ;
ROHLMANN, A .
JOURNAL OF BIOMECHANICS, 1993, 26 (08) :969-990
[3]
BIZOT P, 2000, CLIN ORTHOP RELAT R, V379, P68
[4]
THE USE OF DENSE ALUMINA ALUMINA CERAMIC COMBINATION IN TOTAL HIP-REPLACEMENT [J].
BOUTIN, P ;
CHRISTEL, P ;
DORLOT, JM ;
MEUNIER, A ;
DEROQUANCOURT, A ;
BLANQUAERT, D ;
HERMAN, S ;
SEDEL, L ;
WITVOET, J .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1988, 22 (12) :1203-1232
[5]
Risk factors in falls among the elderly according to extrinsic and intrinsic precipitating causes [J].
Bueno-Cavanillas, A ;
Padilla-Ruiz, F ;
Jiménez-Moleón, JJ ;
Peinado-Alonso, CA ;
Gálvez-Vargas, R .
EUROPEAN JOURNAL OF EPIDEMIOLOGY, 2000, 16 (09) :849-859
[6]
AGE-RELATED REDUCTIONS IN THE STRENGTH OF THE FEMUR TESTED IN A FALL-LOADING CONFIGURATION [J].
COURTNEY, AC ;
WACHTEL, EF ;
MYERS, ER ;
HAYES, WC .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1995, 77A (03) :387-395
[7]
Femoral prosthetic heads and their significance in polyethylene wear [J].
Falez, F ;
La Cava, F ;
Panegrossi, G .
INTERNATIONAL ORTHOPAEDICS, 2000, 24 (03) :126-129
[8]
GARINO JP, 2000, CLIN ORTHOP RELAT R, V379, P41
[9]
Hall R M, 1998, J Orthop Sci, V3, P204, DOI 10.1007/s007760050043
[10]
Heros R., 1998, SEMIN ARTHROPLASTY, V9, P114