Determination of bile acids in human serum by on-line restricted access material-ultra high-performance liquid chromatography-mass spectrometry

被引:59
作者
Bentayeb, K. [1 ]
Batlle, R. [1 ]
Sanchez, C. [1 ]
Nerin, C. [1 ]
Domeno, C. [1 ]
机构
[1] Univ Zaragoza, CPS, Dept Analyt Chem, Aragon Inst Engn Res I3A, E-50018 Zaragoza, Spain
来源
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES | 2008年 / 869卷 / 1-2期
关键词
bile acids; on-line method; restricted access material; ultra high-performance liquid; chromatography; loop-type interface; auxiliary flow;
D O I
10.1016/j.jchromb.2008.04.045
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
This paper describes a new, fully automated on-line method combining restricted access material (RAM) extraction and ultra high-performance liquid chromatography (UHPLC) with mass spectrometric (MS) detection for determining congeners of bile acids (BAs) in human serum. In this method, low-pressure RAM and high-pressure UHPLC-MS are hyphenated by using a 2.5-mL loop-type interface. The compatibility problem between the large volume (1.2 mL) of strong solvent (methanol) used for RAM elution and the need for a weak solvent in UHPLC injection has been addressed by using an auxiliary pre-column cross-flow of 0.1% aqueous formic acid. In this way, the complete 2.5 mL loop volume can be injected into the UHPLC system, thereby maximizing sensitivity while maintaining good chromatographic performance. The optimised method allows the simultaneous analysis of 13 bile acids in a single run, including glycine- and taurine-conjugated bile acids, cholic acid (CA), deoxycholic acid (DCA), chenodeoxycholic acid (CDCA), ursodeoxycholic acid (UDCA), and litocholic acid. The complete analysis of a 100-mu L single serum sample is performed in 30 min, providing detection limits in the pg range (corresponding with clinically relevant concentration levels) for all of the analytes except lithocholic acid, intra-day precision values (%R.S.D.) below 4% (except ursodeoxycholic acid) and inter-day precision lower than 15% (except ursodeoxycholic, glycoursodeoxycholic acid (GUDCA) and lithocholic acid). (c) 2008 Elsevier B.V. All rights reserved.
引用
收藏
页码:1 / 8
页数:8
相关论文
共 22 条
[1]   High sensitive analysis of rat serum bile acids by liquid chromatography/electrospray ionization tandem mass spectrometry [J].
Ando, M ;
Kaneko, T ;
Watanabe, R ;
Kikuchi, S ;
Goto, T ;
Iida, T ;
Hishinuma, T ;
Mano, N ;
Goto, J .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2006, 40 (05) :1179-1186
[2]   Simultaneous quantitation of fatty acids, sterols and bile acids in human stool by capillary gas-liquid chromatography [J].
Batta, AK ;
Salen, G ;
Batta, P ;
Tint, GS ;
Alberts, DS ;
Earnest, DL .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2002, 775 (02) :153-161
[3]   Multidimensional on-line solid-phase extraction (SPE) using restricted access materials (RAM) in combination with molecular imprinted polymers (MIP) [J].
Boos, KS ;
Fleischer, CT .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 2001, 371 (01) :16-20
[4]   Differentiated quantification of human bile acids in serum by high-performance liquid chromatography-tandem mass spectrometry [J].
Burkard, I ;
von Eckardstein, A ;
Rentsch, KM .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2005, 826 (1-2) :147-159
[5]   Direct screening of lyophilised biological fluids for bile acids using an evaporative light scattering detector [J].
Criado, A ;
Cárdenas, S ;
Gallego, M ;
Valcárcel, M .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2003, 792 (02) :299-305
[6]   Advances in methodology for the validation of methods according to the International Organization for Standardization Application to the determination of benzoic and sorbic acids in soft drinks by high-performance liquid chromatography [J].
García, I ;
Ortiz, MC ;
Sarabia, L ;
Vilches, C ;
Gredilla, E .
JOURNAL OF CHROMATOGRAPHY A, 2003, 992 (1-2) :11-27
[7]  
Gatti R, 1997, BIOMED CHROMATOGR, V11, P11, DOI 10.1002/(SICI)1099-0801(199701)11:1<11::AID-BMC608>3.0.CO
[8]  
2-2
[9]   LC/ESI-tandem mass spectrometric determination of bile acid 3-sulfates in human urine 3β-sulfooxy-12α-hydroxy-5β-cholanoic acid is an abundant nonamidated sulfate [J].
Goto, Takaaki ;
Myint, Khin Than ;
Sato, Koichi ;
Wada, Osamu ;
Kakiyama, Genta ;
Iida, Takashi ;
Hishinuma, Takanori ;
Mano, Nariyasu ;
Goto, Junichi .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2007, 846 (1-2) :69-77