Accuracy of computer navigation in ear, nose, throat surgery - The influence of matching strategy

被引:43
作者
Claes, J
Koekelkoren, E
Wuyts, FL
Claes, GME
Van den Hauwe, L
Van de Heyning, PH
机构
[1] Univ Antwerp, Univ Antwerp Hosp, Dept Ear Nose Throat & Head & Neck Surg, B-2650 Edegem, Belgium
[2] Univ Antwerp, Univ Antwerp Hosp, Dept Med Imaging, Edegem, Belgium
关键词
D O I
10.1001/archotol.126.12.1462
中图分类号
R76 [耳鼻咽喉科学];
学科分类号
100213 ;
摘要
Objective: To measure the effect of 4 different matching strategies on the accuracy of computer navigation on the face and within the nose and rhinopharynx. Design: Survey. Setting: Laboratory study. Subjects: Six human cadavers studied within 24 hours of death. Interventions: A commercially available navigation system with infrared optical tracking was used for computer navigation on the face and within the nose of the subjects after matching with external fiducials or with 3 different configurations of anatomical landmarks. Navigation errors were measured and correlated to matching strategies and compared through statistical analysis. Results: Matching with external fiducials on the face results in smaller navigation error than matching with anatomical landmarks. The configuration of matching strategies with anatomical landmarks also significantly determines the accuracy of computer navigation, especially when different locations of accuracy measurement are considered. Conclusion: Statistically significant findings have shown that the choice of a matching strategy is a major factor in the accuracy of computer navigation for ear, nose, throat surgery.
引用
收藏
页码:1462 / 1466
页数:5
相关论文
共 15 条
[1]   Computer-aided endoscopic sinus surgery [J].
Anon, JB .
LARYNGOSCOPE, 1998, 108 (07) :949-961
[2]   USE OF A FRAMELESS, ARMLESS STEREOTAXIC WAND FOR BRAIN-TUMOR LOCALIZATION WITH 2-DIMENSIONAL AND 3-DIMENSIONAL NEUROIMAGING [J].
BARNETT, GH ;
KORMOS, DW ;
STEINER, CP ;
WEISENBERGER, J .
NEUROSURGERY, 1993, 33 (04) :674-678
[3]   Intraoperative image-guidance technology [J].
Casiano, RR .
ARCHIVES OF OTOLARYNGOLOGY-HEAD & NECK SURGERY, 1999, 125 (11) :1275-1278
[4]   Image-guided endoscopic surgery: Results of accuracy and performance in a multicenter clinical study using an electromagnetic tracking system [J].
Fried, MP ;
Kleefield, J ;
Gopal, H ;
Reardon, E ;
Ho, BT ;
Kuhn, FA .
LARYNGOSCOPE, 1997, 107 (05) :594-601
[5]   CLINICAL USE OF A FRAMELESS STEREOTAXIC ARM - RESULTS OF 325 CASES [J].
GOLFINOS, JG ;
FITZPATRICK, BC ;
SMITH, LR ;
SPETZLER, RF .
JOURNAL OF NEUROSURGERY, 1995, 83 (02) :197-205
[6]   BrainLab VectorVision neuronavigation system: Technology and clinical experiences in 131 cases [J].
Gumprecht, HK ;
Widenka, DC ;
Lumenta, CB .
NEUROSURGERY, 1999, 44 (01) :97-104
[7]   Computer-aided surgery in the petrous bone [J].
Gunkel, AR ;
Vogele, M ;
Martin, A ;
Bale, RJ ;
Thumfart, WF ;
Freysinger, W .
LARYNGOSCOPE, 1999, 109 (11) :1793-1799
[8]  
Helm P A, 1998, Comput Aided Surg, V3, P51, DOI 10.1002/(SICI)1097-0150(1998)3:2<51::AID-IGS1>3.0.CO
[9]  
2-J
[10]   A COMPARISON OF CONVENTIONAL AND SPIRAL CT - AN EXPERIMENTAL-STUDY ON THE DETECTION OF SPHERICAL LESIONS [J].
KALENDER, WA ;
POLACIN, A ;
SUSS, C .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1994, 18 (02) :167-176