Quantification of nicotine, chlorpyrifos and their metabolites in rat plasma and urine using high-performance liquid chromatography

被引:35
作者
Abu-Qare, AW [1 ]
Abou-Donia, MB [1 ]
机构
[1] Duke Univ, Med Ctr, Dept Pharmacol & Canc Biol, Durham, NC 27710 USA
来源
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES | 2001年 / 757卷 / 02期
关键词
nicotine; chlorpyrifos; nornicotine; cotinine; chlorpyrifos-oxon; 3,5,6-trichloro-2-pyridinol;
D O I
10.1016/S0378-4347(01)00161-X
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
This study describes a high-performance liquid chromatographic method for the separation and quantification of nicotine, its metabolites nornicotine and cotinine, the insecticide chlorpyrifos (O,O-diethyl-O[3,5,6-trichloro-2-pyridinyl] phosphorothioate), and its metabolites chlorpyrifos-oxon (O,O-diethyl-O [3,5,6-trichloro-2-pyridinyl]phosphate), and TCP (3,5,6-trichloro-2-pyridinol) in rat plasma and urine. The compounds were separated using gradient mobile phase of methanol, acetonitrile and water (pH 3.20) at a flow-rate of 0.8 ml/min in a period of 17 min, and gradient UV detection ranging between 260 and 280 nm. The retention times ranged from 3.4 to 16.7 min. The limits of detection were ranged between 20 and 150 ng/ml, while limits of quantitation were 50-200 ng/ml. Average percentage recovery of five spiked plasma samples were 84.7 +/-8.3, 78.2 +/-7.6, 80.1 +/-7.6, 79.0 +/-6.4, 74.0 +/-7.4, 87.6 +/-7.5, and from urine 85.1 +/-5.2, 75.9 +/-7.0, 82.1 +/-6.1, 79.5 +/-6.1, 71.3 +/-7.4 and 81.3 +/-6.9 for nicotine, nornicotine, cotinine chlorpyrifos, chlorpyrifos-oxon and TCP, respectively. Intra-day accuracy and precision for this method were ranged between 2.2-3.6 and 2.1-2.8%, respectively. The relationship between peak areas and concentration was linear over range between 200 and 2000 ng/ml. This method was applied to analyze the above chemicals and metabolites following combined oral administration in rats. (C) 2001 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:295 / 300
页数:6
相关论文
共 30 条
[1]  
ABOUDONIA MB, NEUROTOXICOLOGY, P437
[2]  
Aprea C, 1999, J AOAC INT, V82, P305
[3]   NICOTINE EFFECTS ON EICOSANOID FORMATION AND HEMOSTATIC FUNCTION - COMPARISON OF TRANSDERMAL NICOTINE AND CIGARETTE-SMOKING [J].
BENOWITZ, NL ;
FITZGERALD, GA ;
WILSON, M ;
ZHANG, Q .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1993, 22 (04) :1159-1167
[4]   Comparison of serum and salivary cotinine measurements by a sensitive high-performance liquid chromatography-tandem mass spectrometry method as an indicator of exposure to tobacco smoke among smokers and nonsmokers [J].
Bernert, JT ;
McGuffey, JE ;
Morrison, MA ;
Pirkle, JL .
JOURNAL OF ANALYTICAL TOXICOLOGY, 2000, 24 (05) :333-339
[5]  
Bernert JT, 1997, CLIN CHEM, V43, P2281
[6]  
BYRNE SL, 1998, ENV HLTH PERSPECT, V106, P731
[7]   High-performance liquid chromatographic determination of cotinine in urine in isocratic mode [J].
Ceppa, F ;
El Jahiri, Y ;
Mayaudon, H ;
Dupuy, O ;
Burnat, P .
JOURNAL OF CHROMATOGRAPHY B, 2000, 746 (02) :115-122
[8]   MULTIROUTE EXPOSURE ASSESSMENT AND BIOLOGICAL MONITORING OF URBAN PESTICIDE APPLICATORS DURING STRUCTURAL CONTROL TREATMENTS WITH CHLORPYRIFOS [J].
FENSKE, RA ;
ELKNER, KP .
TOXICOLOGY AND INDUSTRIAL HEALTH, 1990, 6 (3-4) :349-371
[9]   A simple high performance liquid chromatographic method for the quantification of total cotinine, total 3′-hydroxycotinine and caffeine in the plasma of smokers [J].
Ghosheh, OA ;
Browne, D ;
Rogers, T ;
de Leon, J ;
Dwoskin, LP ;
Crooks, PA .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2000, 23 (2-3) :543-549
[10]  
GILMAN AG, 1990, PHARMACOL BASIS THER, P166