An ultra fast electron beam x-ray tomography scanner

被引:121
作者
Fischer, F. [1 ]
Hoppe, D. [1 ]
Schleicher, E. [1 ]
Mattausch, G. [2 ]
Flaske, H. [2 ]
Bartel, R. [2 ]
Hampel, U. [1 ]
机构
[1] Forschungszentrum Dresden Rossendorf eV, Inst Safety Res, D-01314 Dresden, Germany
[2] Fraunhofer Inst Elect Beam & Plasma Technol FEP, D-01277 Dresden, Germany
关键词
electron beam tomography; x-ray tomography; multiphase flow measurement;
D O I
10.1088/0957-0233/19/9/094002
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
This paper introduces the design of an ultra fast x-ray tomography scanner based on electron beam technology. The scanner has been developed for two-phase flow studies where frame rates of 1 kHz and higher are required. Its functional principle is similar to that of the electron beam x-ray CT scanners used in cardiac imaging. Thus, the scanner comprises an electron beam generator with a fast beam deflection unit, a semicircular x-ray production target made of tungsten alloy and a circular x-ray detector consisting of 240 CZT elements with 1.5 mm x 1.5 mm x 1.5 mm size each. The design is optimized with respect to ultra fast imaging of smaller flow vessels, such as pipes or laboratory-scale chemical reactors. In that way, the scanner is capable of scanning flow cross-sections at a speed of a few thousand frames per second which is sufficient to capture flows of a few meters per second velocity.
引用
收藏
页数:11
相关论文
共 11 条
[1]   Evaluation of a limited angle scanned electron beam x-ray CT approach for two-phase pipe flows [J].
Bieberle, M. ;
Hampel, U. .
MEASUREMENT SCIENCE AND TECHNOLOGY, 2006, 17 (08) :2057-2065
[2]   Ultrafast limited-angle-type x-ray tomography [J].
Bieberle, M. ;
Fischer, F. ;
Schleicher, E. ;
Hampel, U. ;
Koch, D. ;
Aktay, K. S. D. C. ;
Menz, H.-J. ;
Mayer, H.-G. .
APPLIED PHYSICS LETTERS, 2007, 91 (12)
[3]   CARDIAC COMPUTED-TOMOGRAPHY [J].
BOYD, DP ;
LIPTON, MJ .
PROCEEDINGS OF THE IEEE, 1983, 71 (03) :298-307
[4]  
GOULD RG, 1992, ULTRAFAST COMPUTED T
[5]   Experimental ultra fast X-ray computed tomography with a linearly scanned electron beam source [J].
Hampel, U ;
Speck, M ;
Koch, D ;
Menz, HJ ;
Mayer, HG ;
Fietz, J ;
Hoppe, D ;
Schleicher, E ;
Zippe, C ;
Prasser, HM .
FLOW MEASUREMENT AND INSTRUMENTATION, 2005, 16 (2-3) :65-72
[6]   DYNAMIC-BIAS IN RADIATION INTERROGATION OF 2-PHASE FLOW [J].
HARMS, AA ;
LARATTA, FAR .
INTERNATIONAL JOURNAL OF HEAT AND MASS TRANSFER, 1973, 16 (07) :1459-1465
[7]  
Hori K., 1993, GRAPHIC SIMULATION V, V21, P96
[8]   A dual sensor flow imaging tomographic system [J].
Johansen, GA ;
Froystein, T ;
Hjertaker, BT ;
Olsen, O .
MEASUREMENT SCIENCE AND TECHNOLOGY, 1996, 7 (03) :297-307
[9]  
Kak A.C. Slaney M., 1999, PRINCIPLES COMPUTERI
[10]  
Misawa M., 2003, Kerntechnik, V68, P85