New TonoPen XL: comparison with the Goldmann tonometer

被引:81
作者
Iester, M
Mermoud, A
Achache, F
Roy, S
机构
[1] Univ Genoa, Dept Neurol & Visual Sci Ophthalmol B, I-16147 Genoa, Italy
[2] Ist Giannina Gaslini, Ophthalm Div, I-16147 Genoa, Italy
[3] Hop Ophthalm Jules Gonin, Lausanne, Switzerland
关键词
glaucoma; Goldmann tonometer; intraocular pressure; tonometer; TonoPen XL;
D O I
10.1038/eye.2001.13
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose To compare the intraocular pressure (IOP) values obtained using a Goldmann tonometer (Haag-Streit) with those obtained with the new Tonopen XL (Mentor), which has certain differences compared with first- and second-generation models. Methods The IOPs of 104 patients were assessed by Goldmann tonometer and Tonopen XL tonometer. Goldmann measurements was done first in 145 eyes and Tonopen measurements were done first in 53 eyes. Four observers measured the IOP. Observers A, B and C used the Goldmann tonometer first and then the Tonopen XL, while observer D used the Tonopen XL first and then the Goldmann tonometer. The results were analysed by descriptive analysis and, when the distribution of the data was normal, paired t-test and Pearson's r coefficient were used to compare and correlate IOP measurements between Goldmann and Tonopen measurements. When the distribution of the data was non-normal, the Wilcoxon matched-pair test and Spearman coefficient were used. The agreement between Goldmann and Tonopen values was also calculated. ANOVA test was used to compare the difference obtained by 'Goldmann minus Tonopen' measurements among the three different observers. Results A statistically significant difference (p < 0.0001) was found between the IOP readings obtained by Goldmann tonometer and the Tonopen XL and a significant correlation was found between the Goldmann values and Tonopen XL values (p < 0.001). When the Goldmann IOP was more than 20 mmHg the Tonopen XL measurements were lower than the Goldmann values. Also in this group this difference was statistically significant. No significant difference was found between Goldmann values and Tonopen values among the three observers, even though a significant difference was found between Goldmann values and Tonopen values for observer B. When the values obtained by first the Goldmann tonometer and then the Tonopen XL were compared with those obtained by first the Tonopen XL and then the Goldmann tonometer, no significant difference was found between the two groups. Conclusion The new Tonopen XL provides similar results to the Goldmann tomometer in 62% of the cases and was slightly less accurate than the Goldmann tonometer for extreme values, just like the previous Tonopen. Nevertheless the precision is good enough for the purpose of adequate screening.
引用
收藏
页码:52 / 58
页数:7
相关论文
共 21 条
[1]   Tono-Pen determination of intraocular pressure in patients with band keratopathy or glued cornea [J].
Azuara-Blanco, A ;
Bhojani, TK ;
Sarhan, AR ;
Pillai, CT ;
Dua, HS .
BRITISH JOURNAL OF OPHTHALMOLOGY, 1998, 82 (06) :634-636
[2]   Latex allergy associated with the latex cover on the TonoPen [J].
Bergwerk, KL ;
Kodsi, SR .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 1999, 127 (01) :91-91
[3]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[4]  
BORDON AF, 1995, J PEDIATR OPHTHALMOL, V32, P373
[5]  
Copt RP, 1999, ARCH OPHTHALMOL-CHIC, V117, P14
[6]   Tonometry in adults and children -: A manometric evaluation of pneumatonometry, applanation, and TonoPen in vitro and in vivo [J].
Eisenberg, DL ;
Sherman, BG ;
McKeown, CA ;
Schuman, JS .
OPHTHALMOLOGY, 1998, 105 (07) :1173-1181
[7]  
FRENKEL REP, 1988, ARCH OPHTHALMOL-CHIC, V106, P750
[8]   TONO-PEN, A NEW TONOMETER [J].
HESSEMER, V ;
ROSSLER, R ;
JACOBI, KW .
INTERNATIONAL OPHTHALMOLOGY, 1989, 13 (1-2) :51-56
[9]   COMPARISON OF INTRAOCULAR-PRESSURE MEASUREMENTS WITH THE OCULAB TONO-PEN VS MANOMETRY IN HUMANS SHORTLY AFTER DEATH [J].
HESSEMER, V ;
ROSSLER, R ;
JACOBI, KW .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 1988, 105 (06) :678-682
[10]   Frequency doubling technique in patients with ocular hypertension and glaucoma -: Correlation with octopus perimeter indices [J].
Iester, M ;
Mermoud, A ;
Schnyder, C .
OPHTHALMOLOGY, 2000, 107 (02) :288-294