共 19 条
The potential of a modified rf glow discharge source for enhanced elemental speciation studies
被引:13
作者:
Belkin, M
[1
]
Waggoner, JW
[1
]
Caruso, JA
[1
]
机构:
[1] Univ Cincinnati, Dept Chem, Cincinnati, OH 45221 USA
来源:
ANALYTICAL COMMUNICATIONS
|
1998年
/
35卷
/
09期
关键词:
D O I:
10.1039/a805387a
中图分类号:
O65 [分析化学];
学科分类号:
070302 ;
081704 ;
摘要:
Radio frequency glow discharge source mass spectrometry (rf GDMS) was employed for determination of tetramethyltin (TMT) with axial sample introduction via capillary gas chromatography (GC). The axial cell design provided homogenous sampling independent of the transfer line capillary position by eliminating discharge asymmetry. Sensitivity value for TMT of 2237 counts pg-l was achieved with the current setup, a nearly three-fold improvement from a previous study (681 counts pg(-1)). The Sn-120(+) ion signal increased with pressure and decreased with rf power. TMT mass spectra with fragment ion signals comparable in intensity to the Sn-120(+) ion signal were obtained. The ion signal ratio (120Sn+ ion intensity/fragment ion intensity) was monitored as a function of expansion stage pressure of the mass spectrometer to access the extent of molecular fragmentation with variable plasma conditions.
引用
收藏
页码:281 / 283
页数:3
相关论文