Indirect identification and compensation of lateral scanner resonances in atomic force microscopes

被引:12
作者
Burns, D. J. [1 ]
Youcef-Toumi, K. [1 ]
Fantner, G. E. [2 ]
机构
[1] MIT, Dept Mech Engn, Mechatron Lab, Cambridge, MA 02139 USA
[2] Ecole Polytech Fed Lausanne, Lab Bio & Nanoinstrumentat, CH-1015 Lausanne, Switzerland
关键词
BIOLOGICAL APPLICATIONS; AFM;
D O I
10.1088/0957-4484/22/31/315701
中图分类号
TB3 [工程材料学];
学科分类号
082905 [生物质能源与材料];
摘要
Improving the imaging speed of atomic force microscopy (AFM) requires accurate nanopositioning at high speeds. However, high speed operation excites resonances in the AFM's mechanical scanner that can distort the image, and therefore typical users of commercial AFMs elect to operate microscopes at speeds below which scanner resonances are observed. Although traditional robust feedforward controllers and input shaping have proven effective at minimizing the influence of scanner distortions, the lack of direct measurement and use of model-based controllers have required disassembling the microscope to access lateral scanner motion with external sensors in order to perform a full system identification experiment, which places excessive demands on routine microscope operators. Further, since the lightly damped instrument dynamics often change from experiment to experiment, model-based controllers designed from offline system identification experiments must trade off high speed performance for robustness to modeling errors. This work represents a new way to automatically characterize the lateral scanner dynamics without addition of lateral sensors, and shape the commanded input signals in such a way that disturbing dynamics are not excited. Scanner coupling between the lateral and out-of-plane directions is exploited and used to build a minimal model of the scanner that is also sufficient to describe the nature of the distorting resonances. This model informs the design of an online input shaper used to suppress spectral components of the high speed command signals. The method presented is distinct from alternative approaches in that neither an information-complete system identification experiment nor microscope modification are required. Because the system identification is performed online immediately before imaging, no tradeoff of performance is required. This approach has enabled an increase in the scan rates of unmodified commercial AFMs from 1-4 lines s(-1) to over 40 lines s(-1).
引用
收藏
页数:14
相关论文
共 35 条
[1]
Allen S, 2003, BIOCHEM SOC T, V31, P1052
[2]
A high-speed atomic force microscope for studying biological macromolecules [J].
Ando, T ;
Kodera, N ;
Takai, E ;
Maruyama, D ;
Saito, K ;
Toda, A .
PROCEEDINGS OF THE NATIONAL ACADEMY OF SCIENCES OF THE UNITED STATES OF AMERICA, 2001, 98 (22) :12468-12472
[3]
[Anonymous], 1999, SYSTEM IDENTIFICATIO
[4]
ATOMIC FORCE MICROSCOPE [J].
BINNIG, G ;
QUATE, CF ;
GERBER, C .
PHYSICAL REVIEW LETTERS, 1986, 56 (09) :930-933
[5]
High-speed atomic force microscopy for large scan sizes using small cantilevers [J].
Braunsmann, Christoph ;
Schaeffer, Tilman E. .
NANOTECHNOLOGY, 2010, 21 (22)
[6]
BURNS DJ, 2006, 4 INT S NAN
[7]
BURNS DJ, 2004, THESIS MIT
[8]
BURNS DJ, 2008, 5 INT WORKSH SIGN PR
[9]
SCAN SPEED LIMIT IN ATOMIC FORCE MICROSCOPY [J].
BUTT, HJ ;
SIEDLE, P ;
SEIFERT, K ;
FENDLER, K ;
SEEGER, T ;
BAMBERG, E ;
WEISENHORN, AL ;
GOLDIE, K ;
ENGEL, A .
JOURNAL OF MICROSCOPY-OXFORD, 1993, 169 :75-84
[10]
A Review of Feedforward Control Approaches in Nanopositioning for High-Speed SPM [J].
Clayton, Garrett M. ;
Tien, Szuchi ;
Leang, Kam K. ;
Zou, Qingze ;
Devasia, Santosh .
JOURNAL OF DYNAMIC SYSTEMS MEASUREMENT AND CONTROL-TRANSACTIONS OF THE ASME, 2009, 131 (06) :1-19