Physiological noise in oxygenation-sensitive magnetic resonance imaging

被引:492
作者
Krüger, G [1 ]
Glover, GH [1 ]
机构
[1] Stanford Univ, Dept Radiol, Lucas MRS Ctr, Palo Alto, CA 94304 USA
关键词
neuroimaging; spiral scan; magnetic field strength; SNR; physiological noise;
D O I
10.1002/mrm.1240
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The physiological noise in the resting brain, which arises from fluctuations in metabolic-linked brain physiology and subtle brain pulsations, was investigated in six healthy volunteers using oxygenation-sensitive dual-echo spiral MRI at 3.0 T. In contrast to the system and thermal noise, the physiological noise demonstrates a signal strength dependency and, unique to the metabolic-linked noise, an echo-time dependency. Variations of the MR signal strength by changing the flip angle and echo time allowed separation of the different noise components and revealed that the physiological noise at 3.0 T (1) exceeds other noise sources and (2) is significantly greater in cortical gray matter than in white matter regions. The SNR in oxygenation-sensitive MRI is predicted to saturate at higher fields, suggesting that noise measurements of the resting brain at 3.0 T and higher may provide a sensitive probe of functional information. (C) 2001 Wiley-Liss, Inc.
引用
收藏
页码:631 / 637
页数:7
相关论文
共 26 条
  • [1] Bandettini P, 1994, P SOC MAGN RESON, V1, P434
  • [2] FUNCTIONAL CONNECTIVITY IN THE MOTOR CORTEX OF RESTING HUMAN BRAIN USING ECHO-PLANAR MRI
    BISWAL, B
    YETKIN, FZ
    HAUGHTON, VM
    HYDE, JS
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1995, 34 (04) : 537 - 541
  • [3] THE INTRINSIC SIGNAL-TO-NOISE RATIO IN NMR IMAGING
    EDELSTEIN, WA
    GLOVER, GH
    HARDY, CJ
    REDINGTON, RW
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1986, 3 (04) : 604 - 618
  • [4] MR CONTRAST DUE TO MICROSCOPICALLY HETEROGENEOUS MAGNETIC-SUSCEPTIBILITY - NUMERICAL SIMULATIONS AND APPLICATIONS TO CEREBRAL PHYSIOLOGY
    FISEL, CR
    ACKERMAN, JL
    BUXTON, RB
    GARRIDO, L
    BELLIVEAU, JW
    ROSEN, BR
    BRADY, TJ
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1991, 17 (02) : 336 - 347
  • [5] Integrated RF coil with stabilization for fMRI human cortex
    Fitzsimmons, JR
    Scott, JD
    Peterson, DM
    Wolverton, BL
    Webster, CS
    Lang, PJ
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1997, 38 (01) : 15 - 18
  • [6] Experimental determination of the BOLD field strength dependence in vessels and tissue
    Gati, JS
    Menon, RS
    Ugurbil, K
    Rutt, BK
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1997, 38 (02) : 296 - 302
  • [7] Self-navigated spiral fMRI: Interleaved versus single-shot
    Glover, GH
    Lai, S
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1998, 39 (03) : 361 - 368
  • [8] MOTION ARTIFACTS IN FMRI - COMPARISON OF 2DFT WITH PR AND SPIRAL SCAN METHODS
    GLOVER, GH
    LEE, AT
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1995, 33 (05) : 624 - 635
  • [9] Glover GH, 1999, MAGNET RESON MED, V42, P412, DOI 10.1002/(SICI)1522-2594(199908)42:2<412::AID-MRM25>3.0.CO
  • [10] 2-U