Accuracy of five on-site immunoassay drugs-of-abuse testing devices

被引:25
作者
Taylor, EH
Oertli, EH
Wolfgang, JW
Mueller, E
机构
[1] Natl Safety Alliance, Nashville, TN 37211 USA
[2] Lab Specialists Inc, Gretna, LA 70053 USA
关键词
D O I
10.1093/jat/23.2.119
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Many current 'on-site' urine drug-testing products claim performance equivalent to laboratory testing. Five commercially available products (PharmScreen(TM), Roche TestCup®, Accusign(TM) DOA 2, Status DS(TM), and American Bio Medica-Rapid Drug Screen) were challenged with quality-control specimens of known drug metabolite concentrations, 25% above and 25% below the SAMHSA cutoffs, and with known positive and negative donor specimens previously analyzed by immunoassay and gas chromatography-mass spectrometry. The results indicate discrepancies between claims and performance for all products, particularly with amphetamines. The implications for employer- based drug testing are discussed.
引用
收藏
页码:119 / 124
页数:6
相关论文
共 17 条
[11]  
*PRINC BIOM, 1996, ACC DOA 2 PACK INS R
[12]  
*ROCH DIAGN SYST, 1997, ROCH TESTC PACK INS
[13]   APPLICATION OF THE NOVEL IMMUNOASSAY TRIAGE(TM) TO A RAPID DETECTION OF ANTEMORTEM DRUG-ABUSE [J].
ROHRICH, J ;
SCHMIDT, K ;
BRATZKE, H .
JOURNAL OF ANALYTICAL TOXICOLOGY, 1994, 18 (07) :407-414
[14]   ONTRAK TESTCUP - A NOVEL, ON-SITE, MULTIANALYTE SCREEN FOR THE DETECTION OF ABUSED DRUGS [J].
TOWT, J ;
TSAI, SCJ ;
HERNANDEZ, MR ;
KLIMOV, AD ;
KRAVEC, CV ;
ROUSE, SL ;
SUBUHI, HS ;
TWAROWSKA, B ;
SALAMONE, SJ .
JOURNAL OF ANALYTICAL TOXICOLOGY, 1995, 19 (06) :504-510
[15]   EVALUATION OF THE TRIAGE SYSTEM FOR EMERGENCY DRUGS-OF-ABUSE TESTING IN URINE [J].
WU, AHB ;
WONG, SS ;
JOHNSON, KG ;
CALLIES, J ;
SHU, DX ;
DUNN, WE ;
WONG, SHY .
JOURNAL OF ANALYTICAL TOXICOLOGY, 1993, 17 (04) :241-245
[16]  
1997, FED REG, V62, P32360
[17]  
1994, FED REG, V59, P29903