Flat-panel display (LCD) versus high-resolution gray-scale display (CRT) for chest radiography: An observer preference study

被引:31
作者
Balassy, C
Prokop, M
Weber, M
Sailer, J
Herold, CJ
Schaefer-Prokop, C
机构
[1] Univ Vienna, Dept Radiol, A-1090 Vienna, Austria
[2] UMC, Dept Radiol, Utrecht, Netherlands
关键词
D O I
10.2214/ajr.184.3.01840752
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 [临床医学]; 100207 [影像医学与核医学]; 1009 [特种医学];
摘要
OBJECTIVE. Our objective was to compare cathode ray tube (CRT) display with liquid crystal display (LCD) for soft-copy viewing of chest radiographs in a clinical setting. MATERIALS AND METHODS. We displayed 80 posteroanterior digital chest radiographs side by side on a 5-megapixel CRT display and a 3-megapixel LCD. Gradation characteristics of both monitors were adjusted to DICOM display standards. Using a 4-point scale, seven radiologists ranked overall image quality and visibility of anatomic landmarks. Data analysis included Wilcoxon's rank sum test to assess the significance of preference for the different display modes and calculation of the percentage of images ranked equally by at least five of the seven radiologists. RESULTS. Wilcoxon's rank sum test found significant preferences (p < 0.001) for the CRT display for visualization of structures in low-attenuation areas of the thorax and for the LCD for visualization of structures in high-attenuation areas of the thorax. Overall image quality was ranked equal by at least five radiologists in 70% of cases, whereas for the remaining images a significant preference was found for the CRT display. CONCLUSION. We conclude that, under subdued ambient lighting conditions and without use of windowing, for most images the overall quality is equal with high-resolution CRT display and LCD. In images judged preferentially, we found a significant superiority for LCD for delineating mediastinal structures and for CRT display for delineating structures in the lung.
引用
收藏
页码:752 / 756
页数:5
相关论文
共 25 条
[1]
Accurate small-spot luminance measurements [J].
Badano, A ;
Flynn, MJ ;
Kanicki, J .
DISPLAYS, 2002, 23 (04) :177-182
[2]
BARTEN PGJ, 1992, P SOC PHOTO-OPT INS, V1666, P57, DOI 10.1117/12.135956
[3]
Computed radiography in neonatal and pediatric intensive care units: A comparison of 2.5 K x 2 K soft-copy images vs digital hard-copy film [J].
Brill, PW ;
Winchester, P ;
Cahill, P ;
Lesser, M ;
Durfee, SM ;
Giess, CS ;
Auld, PAM ;
Greenwald, B .
PEDIATRIC RADIOLOGY, 1996, 26 (05) :333-336
[4]
Cohen J., 1988, STAT POWER ANAL BEHA
[5]
DIGITAL IMAGING WITH A PHOTOSTIMULABLE PHOSPHOR IN THE CHEST OF NEWBORNS [J].
COHEN, MD ;
KATZ, BP ;
KALASINSKI, LA ;
WHITE, SJ ;
SMITH, JA ;
LONG, B .
RADIOLOGY, 1991, 181 (03) :829-832
[6]
VOLUME DETECTION THRESHOLD - QUANTITATIVE COMPARISON OF COMPUTED RADIOGRAPHY AND SCREEN-FILM RADIOGRAPHY IN DETECTION OF PNEUMOTHORACES IN AN ANIMAL-MODEL THAT SIMULATES THE NEONATE [J].
DON, S ;
COHEN, MD ;
KRUGER, RA ;
WINKLER, TA ;
KATZ, BP ;
LI, W ;
DREESEN, RG ;
KENNAN, N ;
TARVER, R ;
KLATTE, EC .
RADIOLOGY, 1995, 194 (03) :727-730
[7]
FAJARDO LL, 1989, AJR, V152, P474
[8]
FUCHS MH, 2001, RADIOLOGY, V217, P549
[9]
DIGITAL AND CONVENTIONAL CHEST IMAGES - OBSERVER PERFORMANCE WITH FILM DIGITAL RADIOGRAPHY SYSTEM [J].
GOODMAN, LR ;
FOLEY, WD ;
WILSON, CR ;
RIMM, AA ;
LAWSON, TL .
RADIOLOGY, 1986, 158 (01) :27-33
[10]
IK versus 2K monitor: A clinical alternative free-response receiver operating characteristic study of observer performance using pulmonary nodules [J].
Graf, B ;
Simon, U ;
Eickmeyer, F ;
Fiedler, V .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2000, 174 (04) :1067-1074