Implementation of an in vivo diode dosimetry program and changes in diode characteristics over a 4-year clinical history

被引:35
作者
Jursinic, PA [1 ]
机构
[1] Med Coll Wisconsin, Dept Radiat Oncol, Milwaukee, WI 53226 USA
关键词
in vivo dosimetry; diodes; patient dose verification; quality assurance; semiconductor detector;
D O I
10.1118/1.1388217
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
An in vivo dosimetry system that used n-type semiconductor diodes with integral build-up caps was introduced into the clinic. Measurements were made on the entrance surface of the patient and were compared to calculated diode readings expected from monitor units delivered by each beam. A method is given for calibration and correction for chan-es in diode sensitivity, dose-per-pulse effects, collimated field-size (head-scatter factor), wedges, compensators, and scatter from blocks and block trays. Clinically relevant temperature corrections are determined based on temperature measurements made with the diode used as a thermistor. Changes in diode characteristics over 4 years of clinical use are presented. With proper correction for clinical variables it is shown that difference between calculated and measured diode readings are within +/-1% for phantom measurements and within +/-3% for clinical measurements at a 95% confidence level. The correlation of dose measurements on the patient surface to dose inside a target volume is discussed. (C) 2001 American Association of Physicists in Medicine.
引用
收藏
页码:1718 / 1726
页数:9
相关论文
共 34 条
[1]   An audit of radiotherapy patient doses measured with in vivo semiconductor detectors [J].
Adeyemi, A ;
Lord, J .
BRITISH JOURNAL OF RADIOLOGY, 1997, 70 (832) :399-408
[2]   A method to improve the effectiveness of diode in vivo dosimetry [J].
Alecu, R ;
Alecu, M ;
Ochran, TG .
MEDICAL PHYSICS, 1998, 25 (05) :746-749
[3]  
Alecu R, 1999, Med Dosim, V24, P5, DOI 10.1016/S0958-3947(98)00045-4
[4]   Dose perturbations due to in vivo dosimetry with diodes [J].
Alecu, R ;
Feldmeier, JJ ;
Alecu, M .
RADIOTHERAPY AND ONCOLOGY, 1997, 42 (03) :289-291
[5]  
*AM ASS PHYS MED, 1993, 40 AM ASS PHYS MED R
[6]   THE RELATION BETWEEN WEDGE FACTORS IN AIR AND WATER [J].
BARDEROMA, RD ;
BJARNGARD, BE .
MEDICAL PHYSICS, 1994, 21 (07) :1043-1047
[7]  
BJARNGARD BE, 1982, MED PHYS, V9, P258, DOI 10.1118/1.595161
[8]   IN-VIVO DOSIMETRY DURING CONFORMAL THERAPY OF PROSTATIC-CANCER [J].
ESSERS, M ;
LANSON, JH ;
MIJNHEER, BJ .
RADIOTHERAPY AND ONCOLOGY, 1993, 29 (02) :271-279
[9]   RADIATION-DAMAGE INDUCED DOSE-RATE NON-LINEARITY IN AN N-TYPE SILICON DETECTOR [J].
GRUSELL, E ;
RIKNER, G .
ACTA RADIOLOGICA ONCOLOGY, 1984, 23 (06) :465-469
[10]   LINEARITY WITH DOSE-RATE OF LOW-RESISTIVITY P-TYPE SILICON SEMICONDUCTOR-DETECTORS [J].
GRUSELL, E ;
RIKNER, G .
PHYSICS IN MEDICINE AND BIOLOGY, 1993, 38 (06) :785-792