Determination of D-fructose in foodstuffs by an improved amperometric biosensor based on a solid binding matrix

被引:20
作者
Stredansky, M
Pizzariello, A
Stredanska, S
Miertus, S
机构
[1] POLYtech, I-34012 Trieste, Italy
[2] Int Ctr Sci & High Technol, I-34012 Trieste, Italy
来源
ANALYTICAL COMMUNICATIONS | 1999年 / 36卷 / 02期
关键词
D O I
10.1039/a900048h
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An improved amperometric biosensor based on a solid binding matrix (SBM) composite transducer has been used for the determination of D-fructose in foodstuffs samples. The enzyme, D-fructose dehydrogenase (EC 1.1.99.11), was incorporated directly into a solid composite transducer containing both 2-hexadecanone as SBM and chemically modified graphite. Hexacyanoferrate(III) was used as a redox mediator and the current variation caused by the presence of D-fructose was measured amperometrically. The electrochemical properties and the characteristics of the composite fructose biosensors are described. The amperometric signals were fast, reproducible and linearly proportional to D-fructose concentrations in the range 50 x 10(-6)-10 x 10(-3) mol l(-1), with a correlation coefficient of 0.999. A set of measurements at +0.20 V versus SCE for 2 x 10(-3) mol l(-1) D-fructose yielded a relative standard deviation for the steady-state current of 2.11%. The use of a chemically modified graphite by a mild oxidation step was shown to improve the biosensor selectivity against anionic interferents such as L-ascorbate. The biosensor proved to be stable for 6 months and the assay of D-fructose by this electrode was not influenced by the presence of sugars or other interferents commonly found in food samples. The biosensor was used for the determination of n-fructose in some food samples, and the results were consistent with those obtained with the commercially available D-fructose enzyme photometric kit.
引用
收藏
页码:57 / 61
页数:5
相关论文
共 34 条
[1]   A SIMPLE AND RAPID COLORIMETRIC METHOD FOR LACTOSE DETERMINATION IN MILK [J].
ABULEHIA, IH .
FOOD CHEMISTRY, 1987, 24 (03) :233-240
[2]   CHARACTERIZATION OF MIXED ENZYME-MEDIATOR-CARBON PASTE ELECTRODES [J].
AMINE, A ;
KAUFFMANN, JM ;
GUILBAULT, GG ;
BACHA, S .
ANALYTICAL LETTERS, 1993, 26 (07) :1281-1299
[3]   KINETIC-BEHAVIOR OF AN AMPEROMETRIC BIOSENSOR MADE OF AN ENZYMATIC CARBON-PASTE AND A NAFION(R) GEL INVESTIGATED BY ROTATING ELECTRODE STUDIES [J].
ANDRIEUX, CP ;
AUDEBERT, P ;
BACCHI, P ;
DIVISIABLOHORN, B .
JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1995, 394 (1-2) :141-148
[4]   A tri-enzyme electrode probe for the sequential determination of fructose and glucose in the same sample [J].
Antiochia, R ;
Palleschi, G .
ANALYTICAL LETTERS, 1997, 30 (04) :683-697
[5]   NEW ELECTROCATALYTIC BIOMOLECULAR INTERFACE FOR FABRICATING A FRUCTOSE DEHYDROGENASE-BASED SENSING SYSTEM [J].
BEGUM, A ;
KOBATAKE, E ;
SUZAWA, T ;
IKARIYAMA, Y ;
AIZAWA, M .
ANALYTICA CHIMICA ACTA, 1993, 280 (01) :31-36
[6]   Control of electron transfer kinetics at glassy carbon electrodes by specific surface modification [J].
Chen, PH ;
McCreery, RL .
ANALYTICAL CHEMISTRY, 1996, 68 (22) :3958-3965
[7]   CARBON-FIBERS AS ELECTRODE MATERIALS FOR THE CONSTRUCTION OF PEROXIDASE-MODIFIED AMPEROMETRIC BIOSENSORS [J].
CSOREGI, E ;
GORTON, L ;
MARKOVARGA, G .
ANALYTICA CHIMICA ACTA, 1993, 273 (1-2) :59-70
[8]   THE EFFECT OF PH ON SOME OUTER-SPHERE ELECTRODE-REACTIONS AT CARBON ELECTRODES [J].
DEAKIN, MR ;
STUTTS, KJ ;
WIGHTMAN, RM .
JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1985, 182 (01) :113-122
[9]  
Frattali V. P., 1982, FOOD CARBOHYDRATES, P37
[10]   A new amperometric biosensor for fructose using a carbon paste electrode modified with silica gel coated with Meldola's Blue and fructose 5-dehydrogenase [J].
Garcia, CAB ;
Neto, GD ;
Kubota, LT ;
Grandin, LA .
JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1996, 418 (1-2) :147-151