Microfabricated on-chip-type electrochemical flow immunoassay system for the detection of histamine released in whole blood samples

被引:62
作者
Lim, TK [1 ]
Ohta, H [1 ]
Matsunaga, T [1 ]
机构
[1] Tokyo Univ Agr & Technol, Dept Biotechnol & Life Sci, Koganei, Tokyo 1848588, Japan
关键词
D O I
10.1021/ac020749n
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
This paper describes an on-chip-type electrochemical flow immunoassay system with a multichanneled matrix column. The multichanneled matrix column was functionally coated with cation-exchange resin and used for separation of proteins. Antihistamine immunoglobulin G (IgG) antibody conjugated with ferrocenemonocarboxylic acid (Fc) was also prepared and used as a novel analytical reagent. Antibody-antigen complexes were separated from free Fc-conjugated IgG antibody (Fc-IgG) on the basis of differences in isoelectric point (pI) using the multichanneled matrix column coated with cation-exchange resin. The assay yields a good relationship between current and histamine concentration in the range of 200-2000 ng/ mL. This simple technique enables the assay of histamine released in whole blood within 2 min. Furthermore, a good correlation was found between the response of the electrochemical immunoassay described in this paper and the conventional RIA (radioimmunoassay). This on-chip-type electrochemical flow immunoassay requires only minute quantities of whole blood samples and generates highly reproducible results.
引用
收藏
页码:3316 / 3321
页数:6
相关论文
共 46 条
[1]   Protein microchips: Use for immunoassay and enzymatic reactions [J].
Arenkov, P ;
Kukhtin, A ;
Gemmell, A ;
Voloshchuk, S ;
Chupeeva, V ;
Mirzabekov, A .
ANALYTICAL BIOCHEMISTRY, 2000, 278 (02) :123-131
[2]   Passive sensitization of human airways induces mast cell degranulation and release of tryptase [J].
Berger, P ;
N'guyen, C ;
Buckley, M ;
Scotto-Gomez, E ;
Marthan, R ;
Tunon-de-Lara, JM .
ALLERGY, 2002, 57 (07) :592-599
[3]   Analytic chemistry: Everyone's a (future) chemist [J].
Burns, MA .
SCIENCE, 2002, 296 (5574) :1818-1819
[4]   ASSAY OF PROTEIN-QUINONE COUPLING INVOLVING COMPOUNDS STRUCTURALLY RELATED TO ACTIVE PRINCIPLE OF POISON IVY [J].
BYCK, JS ;
DAWSON, CR .
ANALYTICAL BIOCHEMISTRY, 1968, 25 (1-3) :123-&
[5]   Microfabricated devices in biotechnology and biochemical processing [J].
Chován, T ;
Guttman, A .
TRENDS IN BIOTECHNOLOGY, 2002, 20 (03) :116-122
[6]   Electron transfer between ferrocene-modified Au/octadecanethiol/lipid BLM electrode and redox couples in solution [J].
Cui, XL ;
Jiang, DL ;
Diao, P ;
Li, JX ;
Tong, RT ;
Wang, XK .
BIOELECTROCHEMISTRY AND BIOENERGETICS, 1999, 48 (01) :243-247
[7]   ZERO-LENGTH CROSSLINKING PROCEDURE WITH THE USE OF ACTIVE ESTERS [J].
GRABAREK, Z ;
GERGELY, J .
ANALYTICAL BIOCHEMISTRY, 1990, 185 (01) :131-135
[8]   Fabrication of nanocolumns for liquid chromatography [J].
He, B ;
Tait, N ;
Regnier, F .
ANALYTICAL CHEMISTRY, 1998, 70 (18) :3790-3797
[9]   Initial study of using a laminar fluid diffusion interface for sample preparation in high-performance liquid chromatography [J].
Jandik, P ;
Weigl, BH ;
Kessler, N ;
Cheng, J ;
Morris, CJ ;
Schulte, T ;
Avdalovic, N .
JOURNAL OF CHROMATOGRAPHY A, 2002, 954 (1-2) :33-40
[10]  
Kakuta M, 2001, CHEM REC, V1, P395