Application of mid-infrared transmission spectrometry to the direct determination of glucose in whole blood

被引:60
作者
Vonach, R
Buschmann, J
Falkowski, R
Schindler, R
Lendl, B
Kellner, R
机构
[1] Vienna Tech Univ, Inst Analyt Chem, A-1060 Vienna, Austria
[2] BLM Buschmann Lab & Med Technol R&D, D-81669 Munich, Germany
[3] Tech Univ Munich, Inst Proc Control Comp, D-80290 Munich, Germany
关键词
blood analysis; glucose; FT-IR spectrometry; multivariate calibration; blood glucose concentration; plasma sugars; in vitro calibration;
D O I
10.1366/0003702981944553
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The capability of FT-IR transmission spectrometry was examined for the direct determination of glucose in whole blood without any sample preparation. For these investigations, the in hole blood samples were automatically aspirated by a syringe pump into the transmission cell. Infrared spectra were recorded in the 1500-900 cm(-1) range. Despite the high water background absorption and the complex blood matrix, significant spectral changes due to different glucose concentrations were observed. Chemometric (partial least-squares) models were applied for the determination of glucose. A standard error of calibration of 13.8 mg/dL was obtained by using a partial least-squares calibration model containing five ranks. The residues for an independent test set were less than 15 mg/dL.
引用
收藏
页码:820 / 822
页数:3
相关论文
共 11 条
[1]   MULTIVARIATE DETERMINATION OF GLUCOSE IN WHOLE-BLOOD USING PARTIAL LEAST-SQUARES AND ARTIFICIAL NEURAL NETWORKS BASED ON MIDINFRARED SPECTROSCOPY [J].
BHANDARE, P ;
MENDELSON, Y ;
PEURA, RA ;
JANATSCH, G ;
KRUSEJARRES, JD ;
MARBACH, R ;
HEISE, HM .
APPLIED SPECTROSCOPY, 1993, 47 (08) :1214-1221
[2]   Application of molecular spectroscopy in the mid-infrared region to the determination of glucose and cholesterol in whole blood and in blood serum [J].
Budinova, G ;
Salva, J ;
Volka, K .
APPLIED SPECTROSCOPY, 1997, 51 (05) :631-635
[3]   MULTIVARIATE DETERMINATION OF GLUCOSE IN WHOLE-BLOOD BY ATTENUATED TOTAL REFLECTION INFRARED-SPECTROSCOPY [J].
HEISE, HM ;
MARBACH, R ;
JANATSCH, G ;
KRUSEJARRES, JD .
ANALYTICAL CHEMISTRY, 1989, 61 (18) :2009-2015
[4]   MULTICOMPONENT ASSAY FOR BLOOD SUBSTRATES IN HUMAN PLASMA BY MIDINFRARED SPECTROSCOPY AND ITS EVALUATION FOR CLINICAL ANALYSIS [J].
HEISE, HM ;
MARBACH, R ;
KOSCHINSKY, T ;
GRIES, FA .
APPLIED SPECTROSCOPY, 1994, 48 (01) :85-95
[5]   Comparison of evanescent wave spectroscopy based on silver halide fibres with conventional ATR-IR spectroscopy [J].
Heise, HM ;
Bittner, A ;
Kupper, L ;
Butvina, LN .
JOURNAL OF MOLECULAR STRUCTURE, 1997, 410 :521-525
[6]   Rapid and reliable spectral variable selection for statistical calibrations based on PLS-regression vector choices [J].
Heise, HM ;
Bittner, A .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1997, 359 (01) :93-99
[7]   MULTIVARIATE CALIBRATION FOR ASSAYS IN CLINICAL-CHEMISTRY USING ATTENUATED TOTAL REFLECTION INFRARED-SPECTRA OF HUMAN-BLOOD PLASMA [J].
JANATSCH, G ;
KRUSEJARRES, JD ;
MARBACH, R ;
HEISE, HM .
ANALYTICAL CHEMISTRY, 1989, 61 (18) :2016-2023
[8]   Fourier transform infrared detection in miniaturized total analysis systems for sucrose analysis [J].
Lendl, B ;
Schindler, R ;
Frank, J ;
Kellner, R ;
Drott, J ;
Laurell, T .
ANALYTICAL CHEMISTRY, 1997, 69 (15) :2877-2881
[9]   Modulation of the pH in the determination of phosphate with flow injection and Fourier transform infrared detection [J].
Vonach, R ;
Lendl, B ;
Kellner, R .
ANALYST, 1997, 122 (06) :525-530
[10]   Comparison of different infrared measurement techniques in the clinical analysis of biofluids [J].
Wang, J ;
Sowa, M ;
Mantsch, HH ;
Bittner, A ;
Heise, HM .
TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 1996, 15 (07) :286-296