A fatal device-device interaction between a wearable automated defibrillator and a unipolar ventricular pacemaker

被引:13
作者
Lapage, Martin J. [1 ]
Canter, Charles E. [1 ]
Rhee, Edward K. [2 ]
机构
[1] Washington Univ, Sch Med, Dept Pediat, Div Pediat Cardiol, St Louis, MO 63110 USA
[2] St Josephs Hosp, Heart & Lung Inst, Phoenix, AZ USA
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 2008年 / 31卷 / 07期
关键词
defibrillation; ICD; electrophysiology; clinical; pacing; pediatrics; VT;
D O I
10.1111/j.1540-8159.2008.01110.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
We report a fatal device-device interaction between a wearable automated defibrillator (WAD; LifeVest - LifeCor, Inc., Pittsburgh, PA, USA) and a unipolar pacemaker that occurred in an 18-year-old patient listed for cardiac transplantation due to his failing Fontan. The patient developed ventricular tachycardia that was initially detected by the WAD. However, large unipolar pacing artifacts and specific WAD arrhythmia detection algorithms caused the WAD to revert to nonrecognition of the arrhythmia, which lead to the patient's death. We identify likely causes of the failure and suggest methods of preventing such occurrences in the future.
引用
收藏
页码:912 / 915
页数:4
相关论文
共 8 条
[1]   Personal security: Clinical applications of the wearable defibrillator [J].
Beauregard, LAM .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2004, 27 (01) :1-3
[2]   CLINICAL INTERACTIONS BETWEEN PACEMAKERS AND AUTOMATIC IMPLANTABLE CARDIOVERTER-DEFIBRILLATORS [J].
CALKINS, H ;
BRINKER, J ;
VELTRI, EP ;
GUARNIERI, T ;
LEVINE, JH .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1990, 16 (03) :666-673
[3]   THE USE AND INTERACTION OF PERMANENT PACEMAKERS AND THE AUTOMATIC IMPLANTABLE CARDIOVERTER DEFIBRILLATOR [J].
COHEN, AI ;
WISH, MH ;
FLETCHER, RD ;
MILLER, FC ;
MCCORMICK, D ;
SHUCK, J ;
SHAPIRA, N ;
DELNEGRO, AA .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1988, 11 (06) :704-711
[4]  
*FDA, 2001, P010030 FDA
[5]   Use of a wearable defibrillator in terminating tachyarrhythmias in patients at high risk for sudden death: Results of WEARIT/BIROAD [J].
Feldman, AM ;
Klein, H ;
Tchou, P ;
Murali, S ;
Hall, WJ ;
Mancini, D ;
Boehmer, J ;
Harvey, M ;
Hellman, MS ;
Szymkiewicz, SJ ;
Moss, AJ .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2004, 27 (01) :4-9
[6]  
Gregoratos G, 2002, ACC AHA NASPE 2002 G
[7]   UNIPOLAR PACER ARTIFACTS INDUCED FAILURE OF AN AUTOMATIC IMPLANTABLE CARDIOVERTER DEFIBRILLATOR TO DETECT VENTRICULAR-FIBRILLATION [J].
KIM, SG ;
FURMAN, S ;
WASPE, LE ;
BRODMAN, R ;
FISHER, JD .
AMERICAN JOURNAL OF CARDIOLOGY, 1986, 57 (10) :880-881
[8]  
*LIF COR INC, LIFEVEST TRUVECTOR D