Comparison of immunoassay and gas chromatography-mass spectrometry for measurement of polycyclic aromatic hydrocarbons in contaminated soil

被引:30
作者
Chuang, JC
Van Emon, JM
Chou, YL
Junod, N
Finegold, JK
Wilson, NK
机构
[1] Battelle Mem Inst, Columbus, OH 43201 USA
[2] US EPA, Natl Exposure Res Lab, Las Vegas, NV 89193 USA
[3] Battelle Mem Inst, Durham, NC 27713 USA
关键词
polycyclic aromatic hydrocarbons; PAHs; soil; GC-MS; enzyme-linked immunosorbent assay; ELISA; immunoassay;
D O I
10.1016/S0003-2670(03)00499-9
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Polycyclic aromatic hydrocarbons (PAHs) are frequently encountered in the environment and may pose health concerns due to their carcinogenicity. A commercial enzyme-linked immunosorbent assay (ELISA), was evaluated as a screening method for monitoring PAHs at contaminated sites. The ELISA was a carcinogenic PAH (C-PAH) RaPID assay testing kit that cross-reacts with several PAHs and utilizes benzo[a]pyrene (BaP) as a calibrator. Soil samples were extracted with 50% acetone in dichloromethane (DCM) for analysis by ELISA and gas chromatography-mass spectrometry (GC-MS). The overall method precision was within +/-30% for ELISA and within +/-20% for GC-MS. Recovery data for spiked soils ranged from 46 to 140% for BaP as determined by ELISA. Recoveries data of the GC-MS surrogate standards, 2-fluorobiphenyl and chrysene, were greater than 70%. The GC-MS procedure detected a total of 19 priority PAHs (2-6-ring PAHs) including seven probable human carcinogens (4-6-ring B2-PAHs). The ELISA results were compared to GC-MS summation results for the total 19 target PAHs as well as for the subset of the seven B2-PAH compounds. For all soil samples, the PAH concentrations derived from ELISA were greater than the sum of B2-PAH concentrations obtained by GC-MS. ELISA determinations were also frequently greater than the results obtained by GC-MS for the total 19 PAH compounds. This discrepancy can be expected, since the ELISA is a screening assay for the detection of several related PAHs while the GC-MS procedure detects priority PAH compounds. Thus, only a subset of PAHs (e.g. 19 PAHs) in the soil samples were measured by GC-MS while additional PAHs, including alkylated PAHs, and PAH derivatives have been demonstrated to be cross-reactive in the C-PAH ELISA. Results of paired tests show that the PAH data from ELISA and GC-MS methods are significantly different (P < 0.001), but highly correlated. The ELISA data had a strong positive relationship with the GC-MS summation data for the B2-PAHs as well as for the 19 PAHs targeted by the GC-MS method. Results indicate that the ELISA may be useful as a broad screen for monitoring PAHs in environmental samples. (C) 2003 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:31 / 39
页数:9
相关论文
共 17 条
[1]  
ABBOTT J, 2001, P 6 INT S IN SIT ON
[2]  
[Anonymous], 1996, 4035 US EPA
[3]   Evaluation of enzyme-linked immunosorbent assay for the determination of polycyclic aromatic hydrocarbons in house dust and residential soil [J].
Chuang, JC ;
Pollard, MA ;
Chou, YL ;
Menton, RG ;
Wilson, NK .
SCIENCE OF THE TOTAL ENVIRONMENT, 1998, 224 (1-3) :189-199
[4]   MONITORING METHODS FOR POLYCYCLIC AROMATIC-HYDROCARBONS AND THEIR DISTRIBUTION IN-HOUSE DUST AND TRACK-IN SOIL [J].
CHUANG, JC ;
CALLAHAN, PJ ;
MENTON, RG ;
GORDON, SM ;
LEWIS, RG ;
WILSON, NK .
ENVIRONMENTAL SCIENCE & TECHNOLOGY, 1995, 29 (02) :494-500
[5]  
Harvey RG, 1991, POLYCYCLIC AROMATIC
[6]  
*IARC, 1985, IARC MON EV CARC RIS, V35, P4
[7]   Quantitative enzyme-linked immunosorbent assay for determination of polychlorinated biphenyls in environmental soil and sediment samples [J].
Johnson, JC ;
VanEmon, JM .
ANALYTICAL CHEMISTRY, 1996, 68 (01) :162-169
[8]   Quantitative ELISA of polychlorinated biphenyls in an oily soil matrix using supercritical fluid extraction [J].
Johnson, JC ;
Van Emon, JM ;
Clarke, AN ;
Wamsley, BN .
ANALYTICA CHIMICA ACTA, 2001, 428 (02) :191-199
[9]   A strategy to validate immunoassay test kits for TNT and PAHs as a field screening method for contaminated sites in Germany [J].
Krämer, PM .
ANALYTICA CHIMICA ACTA, 1998, 376 (01) :3-11
[10]   Quantitative determination of PCBs in soil and water by a magnetic particle-based immunoassay [J].
Lawruk, TS ;
Lachman, CE ;
Jourdan, SW ;
Fleeker, JR ;
Hayes, MC ;
Herzog, DP ;
Rubio, FM .
ENVIRONMENTAL SCIENCE & TECHNOLOGY, 1996, 30 (02) :695-700