Quadrupole time-of-flight versus triple-quadrupole mass spectrometry for the determination of phosphopeptides by precursor ion scanning

被引:113
作者
Steen, H
Küster, B
Mann, M
机构
[1] Odense Univ, Univ So Denmark, Dept Biochem & Mol Biol, Ctr Expt Bioinformat,Prot Interact Lab, DK-5230 Odense M, Denmark
[2] MDS Prote AS, DK-5230 Odense M, Denmark
来源
JOURNAL OF MASS SPECTROMETRY | 2001年 / 36卷 / 07期
关键词
Q(2)-pulsing; phosphotyrosine immonium ion; nanoelectrospray mass spectrometry; precursor ion scanning;
D O I
10.1002/jms.174
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
An API 3000 triple-quadrupole instrument and a QSTAR Pulsar quadrupole time-of-flight (TOF) mass spectrometer were compared for the determination of phosphopeptides by precursor ion scanning in both the positive and negative nanoelectrospray ionization modes. The limits of detection for synthetic phosphopeptides were similar (500 amol mul(-1)) for both types of instruments when monitoring precursors of -79 Da (PO3-). However, the quadrupole TOF system was approximately fivefold more sensitive (1 fmol mul(-1)) than the triple-quadrupole instrument (5 fmol mul(-1)) when monitoring precursors of 216 Da (immonium ion of phosphotyrosine). The recently introduced Q(2)-pulsing function, which enhances the transmission of fragment ions of a selected m/z window from the collision cell into the TOF part, improved the sensitivity of precursor ion scans on a quadrupole TOF instrument. The selectivity of precursor ion scans is much better on quadrupole TOF systems than on triple quadrupoles because the high resolving power of the reflectron-TOF mass analyzer permits high-accuracy fragment ion selection at no expense of sensitivity. This minimizes interferences from other peptide fragment ions (a-, b-, and y- type) of the same nominal mass but with sufficient differences in their exact masses. As a result, the characteristic immonium ion of phosphotyrosine at m/z 216.043 can be utilized for the selective detection of tyrosine phosphorylated peptides. Our data suggest that, in addition to their superior performance for peptide sequencing, quadrupole TOF instruments also offer a very viable alternative to triple quadrupoles for precursor ion scanning, thus combining high sensitivity and selectivity for both MS and MS/MS experiments in one instrument. Copyright (C) 2001 John Wiley & Sons, Ltd.
引用
收藏
页码:782 / 790
页数:9
相关论文
共 40 条
[1]   A multidimensional electrospray MS-based approach to phosphopeptide mapping [J].
Annan, RS ;
Huddleston, MJ ;
Verma, R ;
Deshaies, RJ ;
Carr, SA .
ANALYTICAL CHEMISTRY, 2001, 73 (03) :393-404
[2]  
BATEMAN R, 1998, P 46 ASMS C MASS SPE, P42
[3]   CONTRIBUTIONS OF MASS-SPECTROMETRY TO PEPTIDE AND PROTEIN-STRUCTURE [J].
BIEMANN, K .
BIOMEDICAL AND ENVIRONMENTAL MASS SPECTROMETRY, 1988, 16 (1-12) :99-111
[4]   Preliminary comparison of precursor scans and liquid chromatography-tandem mass spectrometry on a hybrid quadrupole time-of-flight mass spectrometer [J].
Borchers, C ;
Parker, CE ;
Deterding, LJ ;
Tomer, KB .
JOURNAL OF CHROMATOGRAPHY A, 1999, 854 (1-2) :119-130
[5]  
Boyle W J, 1991, Methods Enzymol, V201, P110
[6]   Selective detection and sequencing of phosphopeptides at the femtomole level by mass spectrometry [J].
Carr, SA ;
Huddleston, MJ ;
Annan, RS .
ANALYTICAL BIOCHEMISTRY, 1996, 239 (02) :180-192
[7]   Duty cycle improvement for a quadrupole-time-of-flight mass spectrometer and its use for precursor ion scans [J].
Chernushevich, IV .
EUROPEAN JOURNAL OF MASS SPECTROMETRY, 2000, 6 (06) :471-479
[8]   Orthogonal-injection TOFMS for analyzing biomolecules [J].
Chernushevich, IV ;
Ens, W ;
Standing, KG .
ANALYTICAL CHEMISTRY, 1999, 71 (13) :452A-461A
[9]  
COVEY T, 1991, ADV LIF SCI, P249
[10]   IDENTIFICATION OF PHOSPHORYLATED PEPTIDES FROM COMPLEX-MIXTURES USING NEGATIVE-ION ORIFICE-POTENTIAL STEPPING AND CAPILLARY LIQUID-CHROMATOGRAPHY ELECTROSPRAY-IONIZATION MASS-SPECTROMETRY [J].
DING, JM ;
BURKHART, W ;
KASSEL, DB .
RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 1994, 8 (01) :94-98